GSi CORTITM User Manual Download Page 39

 

   

 

 

Instrument Settings 

 

The  Corti  instrument  allows  the  user  to  change  many  of  the 
instrument's  settings  or  functions.  These  settings  include  wireless 
Device  Pairing,  Clearing  Test  Results,  Auto  Shutdown  Time, 
Minimum  Amplitude  Value,  Save  Mode,  DPOAE  Norms  On/Off, 
Clock Mode, Language, and Reset to Default Settings.  
 
To access the menus to change these functions, press 

CHANGE at 

the main menu (Display 1) and then press 

 SETUP at the Protocol 

Change  (Display  2)  to  enter  the  Clock  menu  (Display  15).  At  the 
Clock menu, 

hold down

 the 

CHANGE key for 3 seconds until the 

READY light (green LED) turns off and release the key. 

 

NOTE:

  Releasing  the  key  will  access  the  menus  to  change  the 

instrument settings. 

 

Wireless Device Pairing 

The  wireless  pairing  menu  (Display  18)  allows  the user  to  pair  the 
Corti unit with a wireless device, such as a thermal printer or personal 
computer, for printing test results and data transfer. 
 
The Corti unit can be paired to only one device; for example either 
the thermal printer or a PC. To establish wireless pairing, turn on the 
device that will be paired with the Corti unit (e.g.: thermal printer). 
Follow the above instructions to access the menu in Display 18 then 
select 

DISCOVER

  to  initiate  discovery  of  available  wireless 

devices.  The  Corti  will  search  for  available  wireless  devices  for 
approximately  15  seconds.  The  Corti  will  display  the  message 
"Please Wait" and the yellow LED will flash.  
 
When  discovery  is  complete,  all  discovered  devices  are  shown 
(Display 19). A compatible thermal printer will appear as "

PRT-##-

##"

 (example: PRT-5c-25) and other devices will be shown by their 

name. Use the 

CHANGE

 buttons to find the desired device and 

then use the 

PAIR button to pair the Corti to the device. 

 
For PC Pairing: On the PC, select Devices and Printers.  Select 

Add 

a Device

. From the list of identified devices, select 

OAE Device

Select and enter the pairing code 

1234.

  Select Next. A device driver 

may be loaded automatically. The first time the Corti Data Manager 
software is launched, select Detect Com Port to finalize the Corti 
and PC wireless connection.  

 

Pairing  will  be  confirmed  (Display  20).  The  pairing  process  is 
complete. Select 

Main Menu to exit the Wireless pairing menu. 

 

NOTE:

  See  the  Troubleshooting  section  on  page  43  if  wireless 

pairing is unsuccessful or if any error messages are displayed. 

 

 

 

                Display 18 

 

                Display 19 

 

               Display 20 

Summary of Contents for CORTITM

Page 1: ...Part Number D 0103241 Rev D Setting The Clinical Standard www grason stadler com Grason Stadler 10395 West 70th Street Eden Prairie Minnesota 55344 800 700 2282 952 278 4402 fax 952 278 4401 e mail in...

Page 2: ...served No part of this publication may be reproduced or transmitted in any form or by any means without the prior written permission of GSI The information in this publication is proprietary to GSI Co...

Page 3: ...Recycling Disposal 11 Regulatory Symbols 12 Introduction 13 How Does the Corti Device Work 14 What are DPOAEs 14 What Are TEOAEs 15 What Do Otoacoustic Emission Results Tell Us 15 What Frequency Range...

Page 4: ...ts 27 Deleting Results 28 Printing to a Thermal Printer 28 Connecting to the Corti Data Manager 29 Understanding Printed Results 30 Understanding the DPOAE Printout 30 Understanding the TEOAE Printout...

Page 5: ...Reset Protocol 42 Save Protocol 42 Cleaning and Maintenance 43 Cleaning and Disinfection 43 Maintenance 44 Probe tube Replacement 44 Troubleshooting 45 Appendix A Specifications 47 Appendix B Flowchar...

Page 6: ...n Distances between Portable and Mobile RF Communications Equipment and the GSI Corti 57 Guidance and Manufacturer s Declaration Electromagnetic Immunity 58 Guidance and Manufacturer s Declaration Ele...

Page 7: ...s of short duration IEC 60645 6 2010 Electroacoustics Audiometric Equipment Part 6 Instruments for the measurement of otoacoustic emissions ISO 14971 2007 Application of Risk Management to Medical Dev...

Page 8: ...into the ear canal Otoacoustic emissions are low level audio frequency sounds that are produced by the cochlea as part of the normal hearing process Available evidence suggests that otoacoustic emiss...

Page 9: ...ler facility If field service is requested there will be no charge for labor or material however there will be a charge for travel expense at the service center s current rate NOTE Changes in the prod...

Page 10: ...en the Corti power supply and an AC outlet No modifications of the equipment are allowed by anyone other than a qualified GSI representative Modification of the equipment could be hazardous The Corti...

Page 11: ...Medical and Dental Equipment Standard Connection of accessories not meeting these requirements may result in electrical leakage currents in excess of those allowed by the standard and present a poten...

Page 12: ...itional tests can be performed Power Low The battery charge level is too low for operation The user must charge the battery before additional tests can be performed Printer Error Indicates a problem w...

Page 13: ...of illumination of the indicator provides a means of identifying the status of the charging function...

Page 14: ...se in the presence of flammable gaseous mixtures Users should consider the possibility of explosions or fire when using this device in close proximity to flammable anesthetic gases Do NOT use the Cort...

Page 15: ...by hearing healthcare professionals i e ENT doctors audiologists and or technicians neonatal nurses and school nurses who have been trained by a hearing healthcare professional Cautions General If the...

Page 16: ...he instrument from the supply mains by removing the micro USB cable from the connector as shown in Figure 2 on page 15 Warning Explosion This system is not explosion proof Do not use in the presence o...

Page 17: ...vices Follow all your respective local laws and regulations for the proper disposal of batteries and any other parts of this system Below is the contact address for proper return or disposal of electr...

Page 18: ...Documents 980 60601 1 Return to Authorized Representative Special Disposal Required 980 60601 1 Reference Number 60601 1 B Patient Applied Part According to IEC60601 1 980 60601 1 Consult Operating In...

Page 19: ...Performance Only those products bearing the UL Classification Mark for the U S and Canada should be considered as being covered by UL s Classification and Follow Up Service and meeting the appropriate...

Page 20: ...pad of the device allow the user to control testing and printing and to reset test protocols The Probe houses the speaker and microphone which produce test stimuli and measure the sound pressure level...

Page 21: ...wed to override other indications that hearing is not normal A full audiologic evaluation should be administered if concerns about hearing sensitivity persist A REFER test result should not be assumed...

Page 22: ...correlates to normal hearing However a passing result using this instrument is not an indication that the full auditory system is normal Thus a PASS result should not be allowed to override other indi...

Page 23: ...mmunication to the Corti Data Manager Charging and connection to the PC can also be conducted directly from the Corti unit The remainder of this manual assumes that charging and communication to the P...

Page 24: ...ion indicates the charging function is in process Fast blink illumination indicates a fault condition in which the user would be directed by the user manual to return the instrument for service During...

Page 25: ...l Use only the eartips approved for use with the instrument The eartips are disposable and must be replaced after each patient After selecting an eartip push it onto the probe tube until it is flush a...

Page 26: ...he point of the probe head Refer to Figure 7 for an example of the probe in the probe holder when the Corti is resting on a table Refer to Figure 8 for example of the probe in the probe holder when pl...

Page 27: ...reen Figure 9 The yellow TEST light will appear briefly just above the display The green READY light will remain on indicating the instrument is ready for use A Flash Screen will appear briefly indica...

Page 28: ...for customizing protocols on the Diagnostic units can be found in the sections Advanced Options for DPOAE Diagnostic Unit on page 35 or Advanced Options for TEOAE Diagnostic Unit on page 38 NOTE The d...

Page 29: ...tube check that the probe connector is fully seated in the socket and try again NOTE To test children with PE tubes the Probe Check needs to be disabled This is accomplished by first inserting the pr...

Page 30: ...e absolute emission levels at each DP test frequency or TE frequency band The shaded area is the Expanded Boys Town Norms template See page 23 for a description of the template Refer to pages 33 and 3...

Page 31: ...own in Table A1 from Gorga M P Neely S T Ohlrich B Hoover B Redner J and Peters J 1997 From laboratory to clinic a large scale study of distortion product otoacoustic emissions in ears with normal hea...

Page 32: ...rotocols the test will pause when noise levels exceed the noise rejection limit Pause is indicated when the Noise Test and Ready lights turn on simultaneously Testing will automatically resume when no...

Page 33: ...efault mode When operating in the Save 250 mode the Corti will save up to 250 tests There are two options in the Save 250 mode 1 The Corti will automatically number each test from 1 to 250 2 The Corti...

Page 34: ...data will be deleted Data can be deleted using the Clear function in the System Menu page 32 Data can be deleted in the Corti from the Data Manager when the device is connected to the Data Manager ca...

Page 35: ...e Micro USB of the cable into the port found at the base of the Corti instrument see Figure 2 For the PC to Corti Cradle connection plug the USB A of the cable into a computer USB port and the USB B c...

Page 36: ...e signal to noise ratio SNR DP level minus the noise floor in dB 14 A P to the right of the SNR if pass criteria were met for that frequency 15 The Value or SNR graph as selected on the Corti 16 MIN i...

Page 37: ...s are rounded up to the nearest integer and are shown as DP 6 NF 0 and SNR 6 This can result in what appears to be an error with regard to the pass criterion If the pass criterion is 6 dB while the ac...

Page 38: ...lay 15 If the time and date are correct press the UP key to escape back to the main menu Changing the Date Time If either the date or time is incorrect press the CHANGE key to access the menu to chang...

Page 39: ...ith the Corti unit e g thermal printer Follow the above instructions to access the menu in Display 18 then select DISCOVER to initiate discovery of available wireless devices The Corti will search for...

Page 40: ...ess NEXT Save Mode Storing Test Results The Corti unit automatically stores only the most recent test result for each ear L R Display 23 but has the capacity to store 250 individual tests The number o...

Page 41: ...teria including the minimum SNR and the number of passing frequencies for overall test Pass The Corti is set with this feature turned OFF when it is shipped from the factory To change the mode to Mini...

Page 42: ...e Language setting Display 28 allows the user to select among several languages To change the language press the CHANGE keys until the desired language is shown Press the NEXT to exit this menu Reset...

Page 43: ...ol To enter the DPOAE Menu 1 Press CHANGE at the main menu Display 1 2 Using the CHANGE buttons select the DPOAE protocol you want to customize the DP 4s protocol is not customizable 3 Press SETUP at...

Page 44: ...ys to select an option and the NEXT key to exit Display 31 NOTE Longer averaging times help to reduce the noise floor which can improve the likelihood of obtaining a passing result particularly with a...

Page 45: ...PASS to 5 when only 4 frequencies are being tested would result in every test being labeled as a REFER NOTE To disable the PASS REFER indication set the number of frequencies for pass to 0 Press the C...

Page 46: ...ing a Test Protocol To enter the TEOAE Menu 1 Press CHANGE at the main menu Display 1 2 Using the CHANGE buttons select the TEOAE protocol you want to customize the TE 64s protocol is not customizable...

Page 47: ...considered a PASS at that frequency The limits for the PASS SNR are 3 dB to 10 dB Pressing the LEFT or RIGHT arrow keys will increase or decrease the requirement This requirement is used in combinati...

Page 48: ...e instrument settings or the settings of any other protocol Save Protocol Once all of the settings have been selected for the protocol these settings can be saved by selecting the SAVE keys Display 0...

Page 49: ...low level disinfection is appropriate for this type of instrument This may not conform to the infection control guidelines of the user s facility The disinfection materials and procedures applied in t...

Page 50: ...ation date Have an authorized service technician perform electrical safety checks on the unit in order to maintain continued compliance to IEC and UL 60601 1 Probe tube Replacement Probe tubes are dis...

Page 51: ...t print Check the printer status Turn the thermal printer on wake from sleep mode by pressing the power button If the printer doesn t turn on plug in the power supply to charge the battery Be sure the...

Page 52: ...probe and retry the test Replace the probe tube Contact GSI for service if problem persists across several patients Fit Error Too Low For a DP test the level of the calibration tone is too low User s...

Page 53: ...nstrument Weight 6 4 oz 180 gm User Interface OLED Display to provide user information and progress of measurement 4 button keypad to control instrument functions Connectors Communications Integrated...

Page 54: ...Appendix B Flowcharts Flowchart Measurement To Start Press DOWN arrow for 1 second Press UP arrow to return to Main Menu...

Page 55: ...Flowchart Setup Menus Press and hold DOWN arrow for three seconds to enter next menu Clock Menu System Menu DP Settings TE Settings...

Page 56: ...mately 31 Hz The NF estimate is obtained by averaging the power in the 4 closest 2 bins to the DP bin FOR TEOAE The calibration phase automatically measures the peak pressure obtained from a sequence...

Page 57: ...ally that the SNR will be greater than 6 dB approximately 7 times out of 100 This is not a limitation of the instrument but a fundamental property of the method used to estimate the SNR in all emissio...

Page 58: ...t hearing loss to 1 or less Note By the binomial distribution two of three frequencies at 8 4 dB or three of six frequencies at 7 dB should also ensure less than 1 probability of passing a moderately...

Page 59: ...d TEOAE measurements in background noise and babble as high as 55 to 65 dB SPL A weighted Briefly explained use of available memory in the Corti processor permits a post hoc statistical analysis that...

Page 60: ...e of Frequ encies F2 Frequency L1 L2 Averaging Time Pass SNR Passing Freq for Test Pass DP 4s 4 2 0 3 0 4 0 5 0 kHz 65 55 4 sec 6 dB 3 DP 2 0 5 0 4 2 0 3 0 4 0 5 0 kHz 65 55 4 sec 6 dB 3 DP 1 5 6 0 6...

Page 61: ...ed not customizable DP Screener DP 2s DP 4s TE Screener TE 32s TE 64s DP and TE Screener Combo DP 2s DP 4s TE 32s TE 64s DPOAE Screening Default Protocols Protocol Name of Frequ encies F2 Frequency L1...

Page 62: ...netic Compatibility Although the instrument fulfils the relevant EMC requirements precautions should be taken to avoid unnecessary exposure to electromagnetic fields e g from mobile phones etc If the...

Page 63: ...ch radiated RF disturbances are controlled The customer or the user of the GSI Corti can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF com...

Page 64: ...erential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial hospital or residential environment Voltage Dips Short Interruptions a...

Page 65: ...xed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked Note...

Reviews: