6.2.10 Performing additional measurements
6.2.11 Selecting a new method
6.2.12 Measure absorbance
Range: -2600 mAbs to +2600 mAbs
Select the desired wavelength from the method list or by entering
the corresponding method number directly.
Note:
To ensure complete reaction times, the user countdown
may be helpful. See section
To perform additional tests using the same
method:
Press the [Test] key.
This display appears.
Confirm with the [Test] key.
or:
Press the [Zero] key to perform a new zero
setting.
This display appears.
Press the [Esc] key to return to the method
selection.
It is also possible to enter the required
method number directly such as [Shift] +
[1][6][0] for CyA-TEST (cyanuric acid).
Confirm your input with the
key.
Method No
Title
900
mAbs 430 nm
910
mAbs 530 nm
920
mAbs 560 nm
930
mAbs 580 nm
940
mAbs 610 nm
950
mAbs 660 nm
The display shows for example the readings
to the left.
Always perform zero setting using a filled
vial, e.g. with deionised water.
This display appears.
Carry out the measurement of the sample.
This display appears.
Test
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Test
Zero
1
6
Esc
0
900 mAbs 430 nm
-2600 mAbs - +2600 mAbs
prepare Zero
press ZERO
500 mAbs
Zero accepted
prepare Test
press TEST