Note:
To cancel the entry, press the [Esc] key.
If you want to display only the results of one code number,
enter the same Code-No twice (as start and end code).
If you want to display all results without code number
(Code-No is 0), enter [Shift] + [0] twice (as start and end code).
Recalling results of one selected method
Deleting stored results
6.3.4 Calibration
Fluoride method 170
Notes:
1. The same batch of
SPADNS reagent
solution must be used
for adjustment and test. If using a new batch of
SPADNS
reagent
solution, a new calibration has to be performed with
the reagent (see Standard Methods 20th, 1998, APHA,
AWWA, WEF 4500 F D., pages 4-82).
2. As the test result is highly dependent on exact sample and
reagent volumes, the sample and reagent volumes should
always be metered very accurately.
Press the [Mode] key.
Press the [Shift] and [3][3] keys.
Confirm with the
key.
This display appears.
Select the required method from the
displayed list or enter the method number
directly.
Confirm with the
key.
In case of differentiated methods, make a
new selection and confirm with the
key.
This display appears.
• To display all stored results of the
selected method, press the
key.
• To print the displayed result, press the
[F3] key.
• To print all selected results, press the
[F2] key.
• Finish with the [Esc] key.
Press the [Mode] key.
Press the [Shift] and [3][4] keys.
Confirm with the
key.
This display appears.
• To retain the data sets in the memory,
press the [Shift] and [0] keys.
• After pressing the [Shift] and [1] keys, the
following acknowledgment is displayed:
To delete, press the
key.
ATTENTION:
All stored test results are deleted!
If you do not want to delete the data, cancel
and exit the menu by pressing the [Esc] key.
Shift
<Storage>
3
>>
Mode
3
<Storage>
Method
80 Bromine
Start:
cancel: Esc
print: F3
print all: F2
20 Acid demand
40 Aluminium T
80 Bromine T
Shift
<Delete data>
4
Delete all data?
Mode
3
<Delete data>
Delete data
Do not delete: Esc
YES: 1, NO: 0
0
1
Press the [Mode] key.
Press the [Shift] and [4][0] keys.
Confirm with the
key.
This display appears.
Press the [Shift] and [3] keys.
This display appears.
1. Fill a clean vial (
∅
24) with exactly 10 ml
of deionised water, close tightly with the
cap.
2. Place the vial in the sample chamber
correctly. See fig.
.
3. Press the [Zero] key.
4. Remove the vial from the sample
chamber.
5. Add exactly 2 ml
SPADNS reagent
solution (No
) to the 10 ml
deionised water. Caution: vial is filled to
the top!
6. Close the vial tightly with the cap and
swirl gently several times to mix the
contents.
7. Place the vial correctly in the sample
chamber. See fig.
.
8. Press the [Test] key.
9. Remove the vial from the sample
chamber, rinse vial and cap thoroughly
and fill the vial with exactly 10 ml
Fluoride standard
(No
,
concentration 1 mg/l F
-
).
10. Add exactly 2 ml
SPADNS reagent
solution (No
) to the 10 ml
Fluoride standard
. Caution: vial is filled
to the top!
11. Place the vial correctly in the sample
chamber. See fig.
.
12. Press the [Test] key.
This display appears.
Confirm with the
key.
Use the [Esc] key to go back to the method
selection.
Select the fluoride method with the [Shift]
and [1][7][0] and
keys.
If an error message appears, please repeat
the calibration.
Shift
<Calibration>
0
1:M191 Ca-Hardness 2T
2:M191 0 Reset 0 cali.
3:M170 Fluoride L
Mode
4
3
<Calibration>
M170 Fluoride
Zero: deionised water
press ZERO
Zero accepted
T1: 0 mg/l F
press TEST
T1 accepted
T2: 1 mg/l F
press TEST
Calibration
accepted
Esc
7
1
0
Error, absorbance
T2>T1