8.13.5. TW Window for Automatic Taring
[ SDO 2500 sub 06 ]
This command defines an amplitude window for the automatic taring. The setting TW = 100 means, that the
system calculates a new tare value, if the averaged net value of the empty scale falls within 100 digits of the net
zero point. The new tare value will be averaged over the time period TI (see below). If the averaged tare value
falls outside this window, then the tare value will not be updated.
Default setting: TW = 0 [= automatic taring disabled]
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
TW
T+00000
Request: TW = 0 d
TW100
OK
Setup: TW = 100 d
8.13.6. TI
Averaging Time for Automatic Taring
[ SDO 2500 sub 07 ]
This command defines the averaging time for the automatic taring. Within this time period the system calculates
an averaged tare value. Default setting: TI = 0 ms.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
TI
T+00000
Request: TI = 0 ms
TI200
OK
Setup: TI = 200 ms
Figure: Time plot of a measurement cycle with the re-trigger function