
GLDM 64.1 Technical Manual rev. 1.30 February
2020
Page 18 of 50
8.2.15. TN Set / Clear Non-Volatile Tare
[ SDO 2300 sub 13 ]
This command sets the tare mode to volatile or non-volatile. Value range is 0 or 1; Factory default is 0 (volatile).
If set to 1 (non-volatile), every set/clear tare will write the value directly to the EEPROM.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
TN
T:000
Actual setting: TN = volatile
TN1
OK
Setup: TN = non-volatile
8.2.16. ZN Set / Clear Non-Volatile Zero
[ SDO 2300 sub 14 ]
This command sets the zero mode to volatile or non-volatile. Value range is 0 or 1; Factory default is 0
(volatile). If set to 1 (non-volatile), every set/clear zero will write the value directly to the EEPROM.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
ZN
Z:000
Actual setting: ZN = volatile
ZN1
OK
Setup: ZN = non-volatile
8.2.17. AZ Absolute zero point calibration (eCal)
[ SDO 2300 sub 02 ]
The command AZ is used as reference point for all weight calculations and will setup in mV/V.
Permitted values are ± 33000 (= ± 3.3000 mV/V).
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
AZ
Z+00796
Request: Zero point @ 0.0796 mV/V
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
AZ_00500
OK
New: Zero point @ 0.0500 mV/V
Factory default
: 00 000d @ 0.0000mV/V input signal.
8.2.18. AG Absolute gain calibration (eCal)
[ SDO 2300 sub 01 ]
The command AG is used as absolute gain (or measuring range) for all weight calculations and will setup in
mV/V. Permitted values are ± 33000 (= ± 3.3000 mV/V).
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
AG
G+010000
Request: gain 10 000d @ 0.1868 mV/V
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
AG_+005000
OK
New: gain 5 000d @ 1.12 mV/V
Factory default
: 20 000d @ 2.0000mV/V input signal.
8.2.19. CS Save the Calibration Data
[ SDO 2004 sub 02 ]
This command results in the calibration data being saved to the EEPROM and causes the TAC to be
incremented by 1.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
CS
OK
Calibration values saved
The CS command saves all of the
calibration group values, as set by CZ, CG, CM‘n’, DS, DP and ZT. The
command returns ERR and has no updating action unless it is proceded by the cmd CE_XXXXX.