GLDM 64.1 Technical Manual rev. 1.30 February
2020
Page 17 of 50
8.2.11. IZ Correction of System Zero
[ n.a. ]
This command can correct the system zero after a successful calibration, e.g. to correct the unknown weight of
a mounting accessory which was used to hold the calibration weight during the calibration procedure. By a
simple parallel shift of the gain curve the sensitivity of the scale will stay unaffected.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
IZ
OK
System zero corrected
8.2.12. ZR Zero Range
[ SDO 2300 sub 11 ]
Sets the zero range manually
– this is the range in increments within which the weighing scale can be zeroed.
Issuing the ZR command without any parameter will return the current value. Permitted values are between the
lower limit of 0 (= factory default setting) and the upper limit of 999999. A value of zero enables the standard
zero range of +/-2% of max.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
ZR100
OK
Setup: Zero range = 100 d
8.2.13. ZI
Initial Zero Range
[ SDO 2300 sub 10 ]
Define the initial zero range (0...999999 d). If ZI is non-zero the device will perform an automatic Set-Zero when
the weight stabilizes with the No-motion settings and the weight is within the ZI range.
Factory default
: 0.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
ZI100
OK
Setup: Initial Zero range = 100 d
8.2.14. TM Tare mode
[ SDO 2300 sub 12 ]
This command sets the tare mode. The tare modes are defined in the table below.
Master (PC / SPS) sends
Slave (GLDM 64.1) responds
Meaning
CE
E+00017 (example)
Request: TAC counter CE17
CE17
OK
Calibration sequence active
TM1
OK
Setup: Tare mode = 1
Tare modes:
TM
Allow tare of negative values
Clear preset tare at return to range 1
0 (Default)
Yes
Yes
1
No
Yes
2
Yes
No
3
No
No
Note
: For OIML R76 compatible applications a tare mode of 1 must be used.