
Page 40
AVFM 5.0 Area-Velocity Flow Meter
RETRIEVE LOG FILE
Plug a USB Flash Memory Drive (not supplied by Greyline) into the USB output
cable from the instrument. The instrument display will show the USB file
download icon until the log file is transferred to the memory card and then display
file download completed icon. The USB flash drive may be removed.
Download file names will appear in this format:
Tag is set according to the
Log Site ID
entered in the instrument Data Logging
menu.
Download letter will be A for the first download from an instrument. B for the
second, then C etc. At the letter Z a - character will appear indicating that the
maximum number of downloads for that instrument are on the USB flash drive.
Older files can be erased or moved from the flash memory drive or a new memory
drive can be used.
OPENING LOG FILES
Install Greyline Logger on your PC or laptop. Refer to the Help menu in the
program for detailed instructions.
Select File/Open/Instrument Log (.log) to open the log file from your USB flash
drive.
Summary of Contents for AVFM 5.0
Page 2: ...Note This page has been left blank intentionally ...
Page 5: ...Page 5 AVFM 5 0 Area Velocity Flow Meter CONNECTIONS ...
Page 7: ...Page 7 AVFM 5 0 Area Velocity Flow Meter CALIBRATION MENU ...
Page 8: ...Page 8 AVFM 5 0 Area Velocity Flow Meter ICONS ...
Page 20: ...Page 20 AVFM 5 0 Area Velocity Flow Meter ...
Page 29: ...Page 29 AVFM 5 0 Area Velocity Flow Meter AREA VELOCITY FLOW DATA SHEET ...
Page 34: ...Page 34 AVFM 5 0 Area Velocity Flow Meter SS PIPE MOUNTING BAND OPTION VSJ ...
Page 36: ...Page 36 AVFM 5 0 Area Velocity Flow Meter ...
Page 44: ...Page 44 AVFM 5 0 Area Velocity Flow Meter Dual Non Contacting Sensor Configuration ...
Page 45: ...Page 45 AVFM 5 0 Area Velocity Flow Meter CONNECTIONS Dual Sensor Configuration ...
Page 51: ......