Page 16
AVFM 5.0 Area-Velocity Flow Meter
RELAY PARAMETERS
Relay
Press
and
or
to select a relay (2 relays are standard, 4
additional are optional).
Function
Press
or
to select
Off
,
Pulse
,
Flow
,
Velocity
or
Level
.
Flow
On
Position the cursor under the numerals and press
or
to set
digits to the relay
On
set point.
Off
set digits to the
Off
set point.
Pulse
Press
and set digits to the flow volume per relay pulse. Use
this feature for remote samplers, chlorinators or totalizers.
Minimum time between pulses is 2.25 seconds and pulse
duration is 350 milliseconds.
Return to
Relay
and enter settings for each relay.
Velocity
On
Position the cursor under the numerals and press
or
to set
digits to the relay
On
set point.
Off
set digits to the
Off
set point.
Level
On
Position the cursor under the numerals and press
or
to set
digits to the relay
On
set point.
Off
set digits to the
Off
set point.
LOE
mode Specify the state of the relay for loss of echo condition:
Off
,
On
or
Hold
.
Press
to return to
Menu Selections
Summary of Contents for AVFM 5.0
Page 2: ...Note This page has been left blank intentionally ...
Page 5: ...Page 5 AVFM 5 0 Area Velocity Flow Meter CONNECTIONS ...
Page 7: ...Page 7 AVFM 5 0 Area Velocity Flow Meter CALIBRATION MENU ...
Page 8: ...Page 8 AVFM 5 0 Area Velocity Flow Meter ICONS ...
Page 20: ...Page 20 AVFM 5 0 Area Velocity Flow Meter ...
Page 29: ...Page 29 AVFM 5 0 Area Velocity Flow Meter AREA VELOCITY FLOW DATA SHEET ...
Page 34: ...Page 34 AVFM 5 0 Area Velocity Flow Meter SS PIPE MOUNTING BAND OPTION VSJ ...
Page 36: ...Page 36 AVFM 5 0 Area Velocity Flow Meter ...
Page 44: ...Page 44 AVFM 5 0 Area Velocity Flow Meter Dual Non Contacting Sensor Configuration ...
Page 45: ...Page 45 AVFM 5 0 Area Velocity Flow Meter CONNECTIONS Dual Sensor Configuration ...
Page 51: ......