SLT32 V2.77
shows software version installed
View Codes? Yes
Select
Yes
and then
È
to display an instrument performance code. Faults
detected by the instrument are stored in the code and will be requested by
Greyline if you call for service or assistance.
Tag
Enter Instrument Tag number (0-9999)
Date MM DD/YYYY
(with optional Data Logger) Press
Æ
and
Ç
or
È
to change Month, Day Year.
Time HH:MM:SS
(with optional Data Logger) Press
Æ
and
Ç
or
È
to change Hours, Minutes,
Seconds
LOE TIME
Press
Æ
and
Ç
or
È
to change the number of seconds without receiving an
echo before the SLT32 displays
ECHO LOSS
, and Control relays change state
as calibrated under Relay Parameters.
Factory default is 30 seconds and is recommended for most applications,
Minimum is 10 seconds and maximum is 60 seconds.
Reset Tot? Yes
(
FLOW
mode only) Position the cursor under
Yes
to reset the totalizer. Store
this selection at the
Store? Yes
prompt.
Temp 25.0C
Indicates current temperature at the sensor head. Press
Æ
to
Min Temp
and
Æ
to
Max Temp
to display the minimum and maximum temperatures the SLT32
has sensed. Press
Æ
to Display °
C
°
F
. Position the cursor under °
C
for celcius
display or °
F
for Farenheight display. Press
È
to return to
Temp
.
ISB?: No
Yes
Select
Yes
only when the instrument is equipped with an optional Intrinsic
Safety Barrier for sensor mounting in hazardous locations. This setting
compensates for the electrical characteristics of an Intrinsic Safety Barrier.
Defaults? Yes
Press
Æ
3 times to select
Yes
. Store to erase all user settings and return
instrument to factory default settings.
Simul
The Output simulation function controls the 4-20mA output, digital display
and control relays. Use it to simplify calibration of remote devices such as
chart recorders or remote displays, and to test Relay set-points. Simulation
digits are shown as a percentage of full scale. Use the
Ç
and
È
Arrow keys to
simulate
0%
(4mA),
100%
(20mA), and increments of .05% (.008mA).
Increments will automatically accelerate when the
Ç
or
È
is continuously
pressed.
4mA ADJ
20mA ADJ
Use to fine tune the 4mA and 20mA calibration of the 4-20mA output.
Adjustment range is +1mA to -1mA in 0.002mA steps
Important: The 4-20mA output will be forced to 4mA and 20mA respectively
during these adjustments.
New Password
position cursor under digits and set new number between
00
and
99
.
Com 24 48 96 192
Optional (with Data Logger) Speed of RS-232C transmission. Select
2400
,
4800
,
9600
or
19200
baud rate (must be set to match baud rate in 'Greyline
Logger' software and any modems used for serial communication).
SLT32 Level & Flow Monitor
Manual Series B.11
Page 17