From
Vol Hrt
or
Flow
press to make your selection. (
Range
or
Level
mode will bypass the
Volume units selection menu).
Vol Hrt
or
Flow
modes will give you the additional choice of
volumetric units:
ft3
-
cubic feet
USG
-
US gallons
USMG
-
US million gallons (
FLOW
only)
IG
-
Imperial gallons
IMG
-
Imperial million gallons (
FLOW
only)
m3
-
cubic meters
L
-
liters
Press
Æ
to position the cursor underneath your selection and press
È
to make your selection.
Vol
or
Hrt
modes will also prompt you to enter a maximum volume. Press
È
from
MxVol
to position
the cursor under the first digit. Use
È
or
Ç
to change the digits or decimal point to enter the actual
maximum volume you will be measuring. Press
Æ
to return to
MxVol
and
È
.
You can enter values up to 6 digits (e.g. 150,000 liters). When the measured tank volume exceeds 4
digits, the SLT32 will automatically display a ‘multiplier’ on the lower alphanumeric display.
FLOW
mode will offer the additional selection of time units:
s
-
seconds
min
-
minutes
hr
-
hours
d
-
day
Note: The SLT32 will display
Er:ILLEGAL UNITS
if your choice of Units exceeds a maximum value
of 9,999,999. Use
US MGD
,
Imp MGD
or
m3/d
so that Units will be 9,999,999 or less. Press
È
to
Store
and
Æ
to
Yes
. Then press
È
or
Ç
to store all your
UNITS/MODE
selections.
SLT32 Level & Flow Monitor
Manual Series B.11
Page 9
UNITS 1000L
= 150,000 Liters