RELAY PARAMETERS
Each relay can be individually calibrated. Press
È
to
R Function
and
Æ
to select:
Off
Æ
Pulse
Æ
Temp
Æ
Level
Æ
Flow
.
Off
mode means the relay will not be energized
Pulse
Appears in Flow mode only. Press
È
and set digits to the flow volume increment required
between relay pulses. Use this feature for remote samplers, chlorinators or totalizers.
Maximum pulse setting is 999,999. Minimum time between pulses is 1 second and pulse
duration is 350 milliseconds.
Temp
References temperature reading from the SLT32's ultrasonic sensor (air temperature at the
sensor location). Press
È
and
Æ
and set the relay On and Off set points.
Level
Range
Volume
Appears in Level, Range, Volume or HRT modes only. Press
È
to
R Function
and
Æ
to
select
Pump
,
LoAlm
(low alarm) or
HiAlm
(high alarm).
Pump
mode
- press
È
and
Æ
and set relay On and Off set points.
LoAlm
mode
- press
È
and set relay On point. Relay will be On with levels below
this point.
HiAlm
mode
- press
È
and set relay On point. Relay will be On with levels above
this point.
LOE
- press
Æ
to select
Off
Æ
On
Æ
Hold
. In a loss of echo condition, the
(Loss Of Echo) relay will turn Off (de-energize), On (energize) or Hold
its current state. Each relay can be individually configured. Press
È
.
To calibrate a relay for echo loss alarm only, set the relay On and Off set points to exactly
the same value, and then set
LOE
to
On
mode. The relay will now energize only when an
echo loss condition occurs.
SLT32 Level & Flow Monitor
Manual Series B.11
Page 14
CALIBRATION
SPECIAL FUNCTION
RELAY PARAMETERS
Pulse
Temp
R1 Function Off
R1on 0.00 ft3
R1on 0.00 C
°
R1
0.00
OFF
°
C
R2 Function Off
Level
Flow
R1 Mode Pump
LoAlm
HiAlm
R1 on 0.00 ft
R1 on 0.00ft3/s
R1 OFF 0.00 ft
R1 LOE Off
On
Pulse
Temp
Level
Flow
Hold
ALT
(1 2 3 4 5 6)