Mounting and operating manual OXY 3610 MP

  

Page 4 von 14 

 

1.5  Reasonably foreseeable misuse 

 

This device must not be used at potentially explosive areas!  
Do not use these products as safety or emergency stop devices or in any other application 
where failure of the product could result in personal injury or material damage. 
Failure to comply with these instructions could result in death or serious injury and material dam-
age.

 

 

This device must not be used at a patient for diagnostic or other medical 
 purpose

 

1.6  Suitability 

This device has been designed and tested in accordance with the safety regulations for electronic devices. However, 
its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety 
advises given in this manual will be adhered to when using the device. 

1.6.1 

During configuration 

 

Caution, acid!   

The electrode contains KOH 

KOH can cause severe chemical burns! 
If there was contact: 
 

• To skin: Flush contacted area with large amounts of water for several minutes. 

 

• To clothing: remove contaminated clothing. 

 

• To eyes: Flush with large amounts of water for at least 15 minutes, obtain medical treatment. 

After swallowing: 
 • Give large volumes of water. DO NOT induce vomiting! 
 • Obtain medical treatment.

 

 

General instructions and safety regulations for electric, light and heavy current plants, including 
domestic safety regulations (e.g. VDE), have to be observed. 

 

Trouble-free operation and reliability of the device can only be guaranteed if the device is not 
subjected to any other climatic conditions than those stated under "Specification". If the device is 
transported from a cold to a warm environment condensation may cause in a failure of the func-
tion. In such a case make sure the device temperature has adjusted to the ambient temperature 
before trying a new start-up. 

 

This device only serves as supervision by the monitoring of essential or other for the customer 
important systems. 
It must not be used instead of compulsory approval monitoring devices and it is not designed for 
that purpose. If this device is used for the monitoring of such systems on its own, the manufac-
turer will not assume liability for damages whatsoever. 

 

 

Summary of Contents for OXY 3610 MP

Page 1: ...3X 6C 07 Operating Manual Dissolved Oxygen Measuring transducer OXY 3610 MP Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Page 2: ...OF THE OXYGEN MEASURING TRANSDUCER 8 7 CONFIGURATION OF THE DEVICE 8 7 1 CONFIGURATION PROCEDURE 8 7 1 1 OFFSET OFFS 8 7 1 2 Average ambient absolute pressure PABS 9 7 1 3 Display damping FiLT 9 7 1 4...

Page 3: ...asured values This manual can be a valuable contribution All instructions have to be well understood and complied with To be sure that there is no risk arising due to misinter pretation of measured va...

Page 4: ...h large amounts of water for several minutes To clothing remove contaminated clothing To eyes Flush with large amounts of water for at least 15 minutes obtain medical treatment After swallowing Give l...

Page 5: ...repair or maintenance With costs The conceptual design particularly when connecting with other devices has to be done thor oughly Internal connections of other devices e g from ground with protective...

Page 6: ...sealing cable fitting b shim c and sealing ring d Pull out the connection socket f Thread cable through parts b c d and into connector housing e To mount the connection cable 2 3 or 4 wire depending...

Page 7: ...normal operation the oxygen content is displayed in the unit mg l By pressing the Key 2 down the temper ature at the electrode can be shown pressing key 3 up shows the state of the electrode in The st...

Page 8: ...k Calibration surrounding Oxygen electrodes are sensitive components Please read the referring manual prior to use During calibration the rating of the electrode is evaluated 100 means perfect state 4...

Page 9: ...8888 The values will be saved The device will start again The display is shown 8888 Bring back the jumper j1 to the parking position after configuration one contact connected the second contact on air...

Page 10: ...uction of the sensor GWO 3600 11 1 1 General The oxygen electrode is an active electrode consisting of a silver cathode and a lead anode using potassium hydrox ide KOH as an electrolyte In case of oxy...

Page 11: ...ements This may also be the reason for errors in the calibration The membrane head is a spare part and can be ordered individually o Filling hole If the electrode is used at high temperatures or if it...

Page 12: ...d vertically from above till the thread catches Then take up head with paper towel and screw onto electrode from underneath Electrolyte will be displaced from the membrane head flowing over Mop up exc...

Page 13: ...2 years at nominal condition Display 1 5 of measured value 0 2 mg l Output signal 0 2 FS Connection 4 20 mA Zweileiter Spannung Dreileiter bzw Vierleiter Add output signal 0 2 FS Auxiliary energy sup...

Page 14: ...x 80 x 55 mm without elbow type plug and sensor sockets Mounting With fixing holes for wall mounting in housing accessible after cover has been removed Mounting distance 50 x 70mm max shaft diameter...

Reviews: