background image

T02.0.3X.6C-07 

Mounting and operating manual OXY 3610 MP

  

Page 2 von 14 

 

Contents 

1

 

SAFETY ..................................................................................................................................................................... 3

 

1.1

 

G

ENERAL NOTE

 ..................................................................................................................................................... 3

 

1.2

 

I

NTENDED USE

 ...................................................................................................................................................... 3

 

1.3

 

Q

UALIFIED STAFF

 .................................................................................................................................................. 3

 

1.4

 

S

AFETY SIGNS AND SYMBOLS

 ................................................................................................................................. 3

 

1.5

 

R

EASONABLY FORESEEABLE MISUSE

 ...................................................................................................................... 4

 

1.6

 

S

UITABILITY

 .......................................................................................................................................................... 4

 

1.6.1

 

During configuration .................................................................................................................. 4

 

1.6.2

 

During normal operation ............................................................................................................ 5

 

2

 

PRODUCT DESCRIPTION ........................................................................................................................................ 5

 

2.1

 

S

COPE OF SUPPLY

................................................................................................................................................. 5

 

2.2

 

T

YPE PLATE

 .......................................................................................................................................................... 5

 

3

 

INSTALLATION AND COMMISSIONING ................................................................................................................. 6

 

3.1

 

G

ENERAL INSTALLATION INSTRUCTIONS

 .................................................................................................................. 6

 

4

 

ASSIGNMENT OF ELBOW-TYPE PLUG ................................................................................................................. 7

 

5

 

DISPLAY FUNCTIONS .............................................................................................................................................. 7

 

6

 

CALIBRATION OF THE OXYGEN MEASURING TRANSDUCER .......................................................................... 8

 

7

 

CONFIGURATION OF THE DEVICE ........................................................................................................................ 8

 

7.1

 

C

ONFIGURATION PROCEDURE

 ................................................................................................................................ 8

 

7.1.1

 

OFFSET “

OFFS

“ .......................................................................................................................... 8

 

7.1.2

 

Average ambient absolute pressure “

PABS

 ............................................................................... 9

 

7.1.3

 

Display damping “

F

i

LT

“ ............................................................................................................... 9

 

7.1.4

 

Save the last parameter values ................................................................................................. 9

 

8

 

ERROR AND SYSTEM MESSAGES ........................................................................................................................ 9

 

9

 

OFFSET SETTINGS FOR DIFFERENT APPLICATIONS ...................................................................................... 10

 

10

 

AMBIENT PRESSURE AS A FUNCTION OF THE ALTITUDE ABOVE SEA LEVEL ....................................... 10

 

11

 

OXYGEN ELECTRODE ....................................................................................................................................... 10

 

11.1

 

C

ONSTRUCTION OF THE SENSOR 

GWO

 

3600 ................................................................................................... 10

 

11.1.1

 

General ................................................................................................................................ 10

 

11.1.2

 

Design .................................................................................................................................. 11

 

11.2

 

W

HEN IS ELECTRODE MAINTENANCE REQUIRED

 ................................................................................................. 11

 

11.3

 

E

LECTRODE MAINTENANCE

 .............................................................................................................................. 12

 

11.4

 

H

OW TO 

O

PERATE

 ........................................................................................................................................... 12

 

12

 

DECOMMISSIONING, RESHIPMENT AND DISPOSAL .................................................................................... 13

 

12.1

 

D

ECOMMISSIONING

 .......................................................................................................................................... 13

 

12.2

 

R

ESHIPMENT AND DISPOSAL

 ............................................................................................................................. 13

 

13

 

SPECIFICATION .................................................................................................................................................. 13

 

 

 

Summary of Contents for OXY 3610 MP

Page 1: ...3X 6C 07 Operating Manual Dissolved Oxygen Measuring transducer OXY 3610 MP Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Page 2: ...OF THE OXYGEN MEASURING TRANSDUCER 8 7 CONFIGURATION OF THE DEVICE 8 7 1 CONFIGURATION PROCEDURE 8 7 1 1 OFFSET OFFS 8 7 1 2 Average ambient absolute pressure PABS 9 7 1 3 Display damping FiLT 9 7 1 4...

Page 3: ...asured values This manual can be a valuable contribution All instructions have to be well understood and complied with To be sure that there is no risk arising due to misinter pretation of measured va...

Page 4: ...h large amounts of water for several minutes To clothing remove contaminated clothing To eyes Flush with large amounts of water for at least 15 minutes obtain medical treatment After swallowing Give l...

Page 5: ...repair or maintenance With costs The conceptual design particularly when connecting with other devices has to be done thor oughly Internal connections of other devices e g from ground with protective...

Page 6: ...sealing cable fitting b shim c and sealing ring d Pull out the connection socket f Thread cable through parts b c d and into connector housing e To mount the connection cable 2 3 or 4 wire depending...

Page 7: ...normal operation the oxygen content is displayed in the unit mg l By pressing the Key 2 down the temper ature at the electrode can be shown pressing key 3 up shows the state of the electrode in The st...

Page 8: ...k Calibration surrounding Oxygen electrodes are sensitive components Please read the referring manual prior to use During calibration the rating of the electrode is evaluated 100 means perfect state 4...

Page 9: ...8888 The values will be saved The device will start again The display is shown 8888 Bring back the jumper j1 to the parking position after configuration one contact connected the second contact on air...

Page 10: ...uction of the sensor GWO 3600 11 1 1 General The oxygen electrode is an active electrode consisting of a silver cathode and a lead anode using potassium hydrox ide KOH as an electrolyte In case of oxy...

Page 11: ...ements This may also be the reason for errors in the calibration The membrane head is a spare part and can be ordered individually o Filling hole If the electrode is used at high temperatures or if it...

Page 12: ...d vertically from above till the thread catches Then take up head with paper towel and screw onto electrode from underneath Electrolyte will be displaced from the membrane head flowing over Mop up exc...

Page 13: ...2 years at nominal condition Display 1 5 of measured value 0 2 mg l Output signal 0 2 FS Connection 4 20 mA Zweileiter Spannung Dreileiter bzw Vierleiter Add output signal 0 2 FS Auxiliary energy sup...

Page 14: ...x 80 x 55 mm without elbow type plug and sensor sockets Mounting With fixing holes for wall mounting in housing accessible after cover has been removed Mounting distance 50 x 70mm max shaft diameter...

Reviews: