H72.0.11.6C-04 

 Operating Manual GMH 3451 

page 5 of 18 

 

 _____________________________________________________ 

_____________________________________________________________________________ 

 

4 Handling 

4.1 Display elements 

 
 
 

 

1

Main display

: conductivity 

(mS/cm, 

µS/cm) 

resistivity (k

cm) 

TDS / total dissolved solids (mg/l) 
salinity (SAL) 

2

Secondary display

:  measuring value temperature 

3 Arrows to selected 

measuring unit

 

4

Warning signal: 

indicates low battery or missing 
calibration 

5

Display elements to show minimum / maximum / 
memorized measuring value 

6

nLF, Lin

:  

display element for selected 

temperature compensation

 

7

%/K, 1/cm:

  

additional configuration units 

8

logg-arrow

:  

logger is ready 

arrow flashing:   automatic recording (Logg CYCL) is 

active 

4.2 Pushbuttons 

 

 

On / Off key  

press shortly:    switch on/off instrument 

 

Set / Menu

press shortly:  change-over between measuring units 

(only  if „InP: SEt“ is selected) 

press for 2 sec. (menu): invoke configuration menu 

 

 

 

 

min/max when taking measurements: 

press shortly:   min. or max. value is displayed 
press for 2 sec:  the corresponding value is deleted  

Configuration

 

to enter values or change settings 

 

CAL

: only at mode ‘cond’=conductivity: 

 
press for 2 sec:  start cell correction adjustment 

 

Store/Quit

Logger off:  

hold and save current measuring value  

  

(‘HLD’ is displayed) 

Logger on: 

Operation of data logger 

– chapter 8 

Set/Menu:  

confirm settings, return to measuring

 

4.3 Connections 

 

Universal output: 

interface, analog output (see 

chapter 9 “Universal output“) 

Permanently connected measuring cell with 
integrated temperature probe  

Power supply: 

 

The mains socket (1,9 mm inner diameter) is located at the left side of the  

instrument for 10,5-12 V DC supply 

 

 

 

All contacts have to be protected against dirt and moisture!

 

MAX DIF

MIN HLD

%

°C

°F

   S

Logg

%/K

AL

mS

nLF

SAL

mg/l

1/cm

Lin

kOhm

1

2

3

4

5

6

ON

OFF

max

Store

CAL

min

Set

Menu

Quit

Summary of Contents for GMH 3451

Page 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Please carefully read these instructions before use Please consider the safety instructions Please keep for future reference as of version 1 3 GMH...

Page 2: ...6 ELECTRODES MEASURING CELLS 8 6 6 1 Design 8 6 6 2 Calibration Adjustment of measuring cells 8 6 7 TEMPERATURE COMPENSATION 8 6 7 1 Temperature compensation nLF according to EN 27888 8 6 7 2 Linear t...

Page 3: ...ficant damage to property at non observance Attention This symbol warns of possible dangers or dangerous situations which can provoke damage to the device or environment at non observance Note This sy...

Page 4: ...ion If and bAt is shown in the lower display the battery has been used up and needs to be replaced However the device will operate correctly for a certain time If bAt is shown in the upper display the...

Page 5: ...shortly change over between measuring units only if InP SEt is selected press for 2 sec menu invoke configuration menu min max when taking measurements press shortly min or max value is displayed pre...

Page 6: ...tached to a belt etc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 D...

Page 7: ...d solids measurement the filtrate dry residue is determined by means of the conductivity and a conversion factor C tdS Well suited for easy concentration measurements of e g salt solutions The determi...

Page 8: ...t see chapter 15 Accuracy Check Adjustment Service 6 7 Temperature compensation The conductivity of aqueous solutions depends on its temperature The temperature dependency is strongly dependent on the...

Page 9: ...Input Selection of measured variable Cond Conductivity rESi Resistivity tdS Total dissolved solids SAL Salinity SEt Change over measured variables by Set key TDS measurement conversion factor only if...

Page 10: ...uld correspond to output 0 V e g for 0 0 S cm 0 0 S cm 400 mS cm Measuring value that should correspond to output 1 V e g for 100 0 mS cm Set Corr Measurement correction Zero point adjustment offset o...

Page 11: ...be used The software also allows easy configuration and starting of the logger When the logger is activated Func Stor or Func CYCL the hold function is no more available the key store is solely used...

Page 12: ...rding Press store for 2 seconds Start menu press again automatic recording is started Each storage process is signalled by shortly displaying St XXXX where XXXX is the number of the saved data set If...

Page 13: ...es In case you want to develop your own software we offer a GMH3000 development package including A universally applicable Windows functions library GMH3000 DLL with documentation that can be used by...

Page 14: ...re different from default settings the device will shortly display Corr after turned on Default setting for offset and scale are off 0 0 i e inputs are not changed Zero point correction Displayed valu...

Page 15: ...orrect pH calibration The device provides the following functions to help with this 12 1 Calibration interval C Int You can input the interval after which the device reminds you to recalibrate The int...

Page 16: ...rer for repair Err 1 Measured value above allowable range Check pressure not within sensor range measuring value to high Sensor defective Return to manufacturer for repair Err 2 Measured value below a...

Page 17: ...at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in...

Page 18: ...l symbols Additional functions Min max hold Adjustment Calibration Cell correction manually or automatically via selectable reference solution Data logger Real time clock Cyclic 10000 data sets cycle...

Reviews: