background image

H72.0.11.6C-04 

 Operating Manual GMH 3451 

page 15 of 18 

 

 _____________________________________________________ 

_____________________________________________________________________________ 

 

Error messages of automatic adjustment/calibration: 

CAL Err.1  Cell correction too high 

Determined cell correction must not exceed 1.2 

CAL Err.2  Cell correction too small  Determined cell correction must not fall below 0.8 
CAL Err.3  Solution of wrong range 

Wrong solution / far beyond tolerance  

CAL Err.4  Wrong temperature 

Beyond permitted temperature:  
0.0  –  34.0 °C (or 0.0 – 27.0 °C at 111.8 mS/cm) 

Alternative to automatic adjustment: 

Manual calculation of cell correction 

with a reference solution  

Example KCl-solution c= 0.01 M: 1413 

μ

S cm

-1 

at 25°C 

At other temperatures switch temperature compensation off (t.Cor = oFF) and use the referring conductivity! 

Conductivity

 displayed

 = 1500 

μ

S cm

-1

 if selected cell correction is 1.000 cm

-1

 (CELL Corr = 1.000) 

Conductivity of solution at solution temperature 25 °C: Conductivity

 real

 = 1413 

μ

S cm

-1

 

Cell correction c = conductivity

 real

 / conductivity

 displayed

 [cm

-1

                         = 1413 / 1970 * cm

-1

                  

= 0.942 cm

-1

 (Enter CELL Corr of 0.942) 

12 GLP 

GLP (Good Laboratory Practice) includes regular check of devices and accessories. For pH measurements it 
is highly important to ensure correct pH calibration. The device provides the following functions to help with 
this. 

12.1  Calibration interval (C.Int) 

You can input the interval after which the device reminds you to recalibrate.  
The interval times should be chosen according to the application and the stability of the electrode. “CAL” 
flashes on the display as soon as the interval has expired. 

12.2  Calibration storage (rEAd CAL)  

The last 16 calibrations are stored with results and date and can be read out. 

Display calibration data: 

Historical calibration data can be comfortably read out via PC software GMHKonfig and GSOFT3050 or 
displayed directly at the device: 

 

 

Press for 2 seconds: 

 

The display will show:

 

 

or

 

(configuration level)

 

 

Press several times until this is 
displayed:

 

 read cal. = “read calibration data” 

 

Press shortly:

 switch between:  

-  CELL = cell correction 
-  C.rEF = reference value, at which cell correction has been adjusted  
-  Display of date+time of data set  

 or 

 

Change between the different calibration data sets 

 

 

Quit calibration data set display  

13 Alarm („AL.“) 

There are 3 possible settings: 
off (AL.oFF), on with buzzer (AL.on), on without buzzer (AL.no.So).  
Alarm is given in the following cases (if alarm active, AL.on or AL.no.So): 

-

  Lower alarm boundary (AL. Lo) under-run  

-

  Upper alarm boundary (AL. Hi) over-rum 

-

 Sensor 

error 

-

 Low 

battery 

(bAt) 

-

  Err.7: system error (always with buzzer!) 

In case of an alarm (and when polling the interface) the ‘PRIO’-flag is set in the returned interface message. 

4

Set

Menu

4

Set

Menu

3

CAL

2

max

5

min

Summary of Contents for GMH 3451

Page 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Please carefully read these instructions before use Please consider the safety instructions Please keep for future reference as of version 1 3 GMH...

Page 2: ...6 ELECTRODES MEASURING CELLS 8 6 6 1 Design 8 6 6 2 Calibration Adjustment of measuring cells 8 6 7 TEMPERATURE COMPENSATION 8 6 7 1 Temperature compensation nLF according to EN 27888 8 6 7 2 Linear t...

Page 3: ...ficant damage to property at non observance Attention This symbol warns of possible dangers or dangerous situations which can provoke damage to the device or environment at non observance Note This sy...

Page 4: ...ion If and bAt is shown in the lower display the battery has been used up and needs to be replaced However the device will operate correctly for a certain time If bAt is shown in the upper display the...

Page 5: ...shortly change over between measuring units only if InP SEt is selected press for 2 sec menu invoke configuration menu min max when taking measurements press shortly min or max value is displayed pre...

Page 6: ...tached to a belt etc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 D...

Page 7: ...d solids measurement the filtrate dry residue is determined by means of the conductivity and a conversion factor C tdS Well suited for easy concentration measurements of e g salt solutions The determi...

Page 8: ...t see chapter 15 Accuracy Check Adjustment Service 6 7 Temperature compensation The conductivity of aqueous solutions depends on its temperature The temperature dependency is strongly dependent on the...

Page 9: ...Input Selection of measured variable Cond Conductivity rESi Resistivity tdS Total dissolved solids SAL Salinity SEt Change over measured variables by Set key TDS measurement conversion factor only if...

Page 10: ...uld correspond to output 0 V e g for 0 0 S cm 0 0 S cm 400 mS cm Measuring value that should correspond to output 1 V e g for 100 0 mS cm Set Corr Measurement correction Zero point adjustment offset o...

Page 11: ...be used The software also allows easy configuration and starting of the logger When the logger is activated Func Stor or Func CYCL the hold function is no more available the key store is solely used...

Page 12: ...rding Press store for 2 seconds Start menu press again automatic recording is started Each storage process is signalled by shortly displaying St XXXX where XXXX is the number of the saved data set If...

Page 13: ...es In case you want to develop your own software we offer a GMH3000 development package including A universally applicable Windows functions library GMH3000 DLL with documentation that can be used by...

Page 14: ...re different from default settings the device will shortly display Corr after turned on Default setting for offset and scale are off 0 0 i e inputs are not changed Zero point correction Displayed valu...

Page 15: ...orrect pH calibration The device provides the following functions to help with this 12 1 Calibration interval C Int You can input the interval after which the device reminds you to recalibrate The int...

Page 16: ...rer for repair Err 1 Measured value above allowable range Check pressure not within sensor range measuring value to high Sensor defective Return to manufacturer for repair Err 2 Measured value below a...

Page 17: ...at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in...

Page 18: ...l symbols Additional functions Min max hold Adjustment Calibration Cell correction manually or automatically via selectable reference solution Data logger Real time clock Cyclic 10000 data sets cycle...

Reviews: