background image

H72.0.11.6C-04 

 Operating Manual GMH 3451 

page 14 of 18 

 

 _____________________________________________________ 

_____________________________________________________________________________ 

 

9.2 Analogue Output 

An analogue voltage 0-1V can be tapped at the universal output socket (mode: “Out dAC”). 
With the DAC.0 and DAC.1 values the output can be rapidly scaled to your efforts. 
Keep in mind not to connect low-resistive loads to the output, otherwise the output value will be wrong and 
battery life is decreased. Loads above ca 10kOhm are uncritical. 
If the display exceeds the value set by DAC.1, then the device will apply 1V to the output 
If the display falls below the value set by DAC.0, then the device will apply 0V to the output   
In case of an error (Err.1, Err.2, no sensor, etc.) the device will apply slightly above 1V to the output. 

 

plug wiring: 

 

GND 
 
+Uout 

Attention! 

the 3

rd

 contact has to be left floating! 

Only stereo plugs are allowed! 

10 Adjustment of temperature input  

The temperature input can be adjusted with offset and scale. A reasonable adjustment presumes reliable 
references (e.g. ice water, controlled precision water bath, etc.). 

If the inputs are adjusted (i.e. offset and scale are different from default settings) the device will shortly 
display “Corr” after turned on.  
Default setting for offset and scale are ‘off’ = 0.0, i.e. inputs are not changed.  

Zero point correction: 

Displayed value = measured value – OFFS  

Zero point and slope correction: 

Displayed value = (measured value – OFFS) * (1 + SCAL / 100)  

Displayed value °F = (meas. value °F - 32°F - OFFS) 

 (1 + SCAL / 100) 

 

11 Automatic adjustment/calibration of cell correction  

Besides the direct input of the cell correction (see below) via the menu (“CELL Corr”) the cell correction can 
also be determined automatically: 

    Manual adjustment

 or 

 

Adjustment with reference solution

  

Menu selection: 

 

 

 “CAL Edit” 

 

 

“CAL rEF.S” 

         Menu 

selection 

of 

desired 

solution 

 

1413 µS/cm  

0.01 M KCL 

2.76 mS/cm 

0.02 M KCL 

12.88 mS/cm   0.1 M KCL 
50 mS/cm  

KCL 

111.8 mS/cm 1 M KCL 

Values for 25°C, the temperature 
dependency of those solutions are known 
by the device and are compensated 
automatically.  

Press Cal-key for 2 s,  
start of calibration 

         “actual value” e.g. “1723 µS/cm” 

“value of solution” e.g. “1413 µS/cm” 

 

 

 

            and CAL with rotary symbol                    and CAL with rotary symbol 

 
Calibration 
 

 

 

 

Select desired display value   

wait until device measures stable value  

 

 

 

           with buttons “up” and “down” 

    and 

confirm 

with 

“enter”

 

 

 
 
Afterwards the device returns to the normal measuring operation mode or – if so – displays an error 
message. 
The resulting cell correction can be seen in the menu at “CELL Corr” and the calibration history.   

 

 

Summary of Contents for GMH 3451

Page 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Please carefully read these instructions before use Please consider the safety instructions Please keep for future reference as of version 1 3 GMH...

Page 2: ...6 ELECTRODES MEASURING CELLS 8 6 6 1 Design 8 6 6 2 Calibration Adjustment of measuring cells 8 6 7 TEMPERATURE COMPENSATION 8 6 7 1 Temperature compensation nLF according to EN 27888 8 6 7 2 Linear t...

Page 3: ...ficant damage to property at non observance Attention This symbol warns of possible dangers or dangerous situations which can provoke damage to the device or environment at non observance Note This sy...

Page 4: ...ion If and bAt is shown in the lower display the battery has been used up and needs to be replaced However the device will operate correctly for a certain time If bAt is shown in the upper display the...

Page 5: ...shortly change over between measuring units only if InP SEt is selected press for 2 sec menu invoke configuration menu min max when taking measurements press shortly min or max value is displayed pre...

Page 6: ...tached to a belt etc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 D...

Page 7: ...d solids measurement the filtrate dry residue is determined by means of the conductivity and a conversion factor C tdS Well suited for easy concentration measurements of e g salt solutions The determi...

Page 8: ...t see chapter 15 Accuracy Check Adjustment Service 6 7 Temperature compensation The conductivity of aqueous solutions depends on its temperature The temperature dependency is strongly dependent on the...

Page 9: ...Input Selection of measured variable Cond Conductivity rESi Resistivity tdS Total dissolved solids SAL Salinity SEt Change over measured variables by Set key TDS measurement conversion factor only if...

Page 10: ...uld correspond to output 0 V e g for 0 0 S cm 0 0 S cm 400 mS cm Measuring value that should correspond to output 1 V e g for 100 0 mS cm Set Corr Measurement correction Zero point adjustment offset o...

Page 11: ...be used The software also allows easy configuration and starting of the logger When the logger is activated Func Stor or Func CYCL the hold function is no more available the key store is solely used...

Page 12: ...rding Press store for 2 seconds Start menu press again automatic recording is started Each storage process is signalled by shortly displaying St XXXX where XXXX is the number of the saved data set If...

Page 13: ...es In case you want to develop your own software we offer a GMH3000 development package including A universally applicable Windows functions library GMH3000 DLL with documentation that can be used by...

Page 14: ...re different from default settings the device will shortly display Corr after turned on Default setting for offset and scale are off 0 0 i e inputs are not changed Zero point correction Displayed valu...

Page 15: ...orrect pH calibration The device provides the following functions to help with this 12 1 Calibration interval C Int You can input the interval after which the device reminds you to recalibrate The int...

Page 16: ...rer for repair Err 1 Measured value above allowable range Check pressure not within sensor range measuring value to high Sensor defective Return to manufacturer for repair Err 2 Measured value below a...

Page 17: ...at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in...

Page 18: ...l symbols Additional functions Min max hold Adjustment Calibration Cell correction manually or automatically via selectable reference solution Data logger Real time clock Cyclic 10000 data sets cycle...

Reviews: