GREISINGER GIA1000 FR Connecting And Operating Manual Download Page 8

GIR1002 FR / GIA1000 FR       Connecting and Operating Manual         Version: 4.03-3       page: 8

Display/setting of breaking point/hysteresis for switching output 1 or 2

To be called up by:

pressing button 1 till LEDs 

SET

, "1" or 

2

 and 

OFF/Hy

 are illuminated.

Display:

The 7-segment display shows the breaking point/hysteresis for switching
output 1 or 2.

Setting:

Set breaking point/hysteresis by means of buttons 2 and 3. LEDs "1" or 

2

and 

ON/Hy

 will start flashing.

Buttons 2 and 3 are equipped with a scrolling function, i.e. the making
point will be increased respectively decreased by 1 digit when button 2
respectively button 3 is pressed briefly (no more than 1 s). When pressing
these buttons for a longer time (over 1 s) the values will start 

scrolling

upwards respectively downwards. After 

scrolling

 of approx. 150 digits the

scrolling speed

 will be increased by a factor of 10.

Use button 4 to acknowledge new breaking point/hysteresis.

LEDs "1" or 

2

 and 

ON/Hy

 will stop flashing.

Switch over to display of current measuring value by pressing button 4 once again.

Display/setting  of  max.  alarm  limit

To be called up by:

pressing button 1 till LEDs 

SET

ALARM

 and  

max.

 are illuminated.

Display:

The 7-segment display shows the max. alarm limit.

Setting:

Set max. alarm limit by means of buttons 2 and 3.

LEDs 

ALARM

 and 

max.

 will start flashing.

Buttons 2 and 3 are equipped with a scrolling function, i.e. the making
point will be increased respectively decreased by 1 digit when button 2
respectively button 3 is pressed briefly (no more than 1 s). When pressing
these buttons for a longer time (over 1 s) the values will start 

scrolling

upwards respectively downwards. After 

scrolling

 of approx. 150 digits the

scrolling speed

 will be increased by a factor of 10.

Use button 4 to acknowledge new max. alarm limit

.

LEDs 

ALARM

 and 

max.

 will stop flashing.

Switch over to display of current measuring value by pressing button 4 once again.

Display/setting  of  min.  alarm  limit

To be called up by:

pressing button 1 till LEDs 

SET

ALARM

 and 

min.

are flashing

Display:

The 7-segment display shows the min. alarm limit.

Setting:

Set min. alarm limit by means of buttons 2 and 3. LEDs 

ALARM

 and

min.

 will start flashing.

Buttons 2 and 3 are equipped with a scrolling function, i.e. the making
point will be increased respectively decreased by 1 digit when button 2
respectively button 3 is pressed briefly (no more than 1 s). When pressing
these buttons for a longer time (over 1 s) the values will start 

scrolling

upwards respectively downwards. After 

scrolling

 of approx. 150 digits the

scrolling speed

 will be increased by a factor of 10.

Use button 4 to acknowledge new min. alarm limit.

LEDs 

ALARM

 and 

min.

 will stop flashing.

Switch over to display of current measuring value by pressing button 4 once again.

Display/setting  of  alarm  delay  [optional]

To be called up by:

Push the button at the rear of the GIR1002 till the LED "SET" is
illuminated. Pressing button 1 till LEDs 

SET

ALARM

 are illuminated.

Display:

The 7-segment display shows "d." and the alarm delay.

Setting:

Set alarm delay by means of buttons 2 and 3.
Use button 4 to acknowledge new min. alarm limit.

Switch over to display of current measuring value by pressing button 4
once again.

Summary of Contents for GIA1000 FR

Page 1: ...rating Manual GIR1002 FR GIA1000 FR Version 4 03 This connecting and operating manual may be subject to technical alternations ON SP SET min OFF Hy 1 2 max GIR GIA ALARM Scale Analog CAL SET 1 2 3 4 I...

Page 2: ...ge 7 Max value reset page 7 Display of min value page 7 Min value reset page 7 Display setting of making point switching point for switching output 1 or 2 page 7 Display setting of breaking point hyst...

Page 3: ...l supply the GIR1002 FR factory tested and completely calibrated Before you can actually use it make sure to configurate your GIR1002 FR for your special application Please also refer to the chapter C...

Page 4: ...connected to a suitable RC element for example RC220 Electric connection and commissioning of the device must be carried out by trained and skilled personnel Wrong connection may lead to the destructi...

Page 5: ...min selectable Interface RS485 Power supply 230VAC or as specified on device housing Power consumption 5W 230VAC 3W 24VDC 1 5W 12VDC Relay switching power 10A 250V AC ohmic load use RC element when sw...

Page 6: ...setting of making point button 1 4 1 2 3 7 display setting of switching point button 1 4 4 5 6 8 display setting of breaking point button 1 4 1 2 3 7 display setting of hysteresis button 1 4 4 5 6 8 S...

Page 7: ...in case of alarm LED min LEDflashing Min value reset To be called up by pressing button 3 for approx 4s Display please refer to display of min value Display setting of makingpoint switching point for...

Page 8: ...briefly no more than 1 s When pressing these buttons for a longer time over 1 s the values will start scrolling upwards respectively downwards After scrolling of approx 150 digits the scrolling speed...

Page 9: ...int to the left or button 3 shift decimal point to the right to set desired decimal point position acknowledge position by pressing button 4 If decimal point is positioned to the right of the lowest d...

Page 10: ...urrent of 20mA LED max will start flashing acknowledge by pressing button 4 8 LEDs Analog min and LED CAL are illuminated The display will show the measuring value stored in the GIR1002 FR for which a...

Page 11: ...rmissibe range This fault code indicates that the min display value of 1999 digits has fallen below Remedies FE 4 will be reset as soon as the values on the display are back within their permissible r...

Page 12: ...This method as well as the fact that the interface operates in the half duplex mode ensure that up to 16 devices can be called via only one 2 wire line Interface configuration Type RS485 Transfer mode...

Page 13: ...s specifying the parameter and or value to be processed For assignments please refer to table 1 of chapter functional code ASCII Code 2FH Indicates the end of a data transfer process Each data transfe...

Page 14: ...g point Switching point 0 9 30H 39H 1 Switching output 2 Breaking point Hysteresis 0 A 30H 41H 1 Max alarm limit 0 B 30H 42H 1 Min alarm limit 0 C 30H 43H 1 Input signal 1 5 31H 35H 3 Position of deci...

Page 15: ...l The assignment is as follows 0 2Vss 1 300mVss 2 30mVss 3 TTL Data format 4 This format is used to call up and or program the position of the decimal point The assignment is as follows 0 no decimal p...

Reviews: