Electric Connection

All connections for the GIR1002 FR are located at the back side of the device.
Connections are made via screw-type/plug-in terminals.

Make it a rule to mount screw-type/plug-in terminals while they are still loose and to put them on only
afterwards. When mounting terminals already put on there is a risk that soldering eyelets may be pulled
out. Please use suitable screw driver and do not tighten screws by force.

Supply Voltage :

230V AC, 50/60Hz or instruction on the device

Please make sure that the mains voltage applied corresponds to the supply voltage stated at the device nameplate

Switching output:

1 potential-free relay (changeover contact)
1 potential-free relay (make-contact)

Switching capacity:

10A, 250V AC (ohmic load)

When using inductive loads the contact has to be connected to a
suitable RC-element (for example 'RC220').

Electric  connection  and  commissioning  of the  device  must be  carried  out by  trained and  skilled
personnel.  Wrong  connection may  lead  to  the  destruction  of the  controller, in  which  case  we
cannot  assume  any  warranty.

Assignment Diagram

Terminal  Assignment

Terminal
number

Max.
terminal
range

Assignment

Notes

1
2

2,5 mm²

Supply voltage
Supply voltage

230 V AC 50/60 Hz
or as specified on device housing

3
4
5

2,5 mm²

Relay, normally-closed
Relay, input
Relay, normally-open

Switching output
or
alarm output

6
7

2,5 mm²

Relay, input
Relay, normally-open

Switching output 1

8
9

1,5 mm²

Transmitter 
Transmitter supply  -

18 V DC / 25mA

10
11
12
13

1,5 mm²

GND

signal

frequency input

14
15

1,5 mm²

Connection B
Connection A

RS485

16
17

1,5 mm²

Analog 
Analog output -

only for devices with optional analog
output

Supply  voltage

8
9

10
11
12
13

3

4

5

6

7

14
15
16
17

1

2

GIR1002FR  (view of  top)

Switching  output  2

Switching  output  1

Configuration key

Frequency  input

RS485

Analog  output    (optional)

Transmitter  supply

Summary of Contents for GIA1000 FR

Page 1: ...rating Manual GIR1002 FR GIA1000 FR Version 4 03 This connecting and operating manual may be subject to technical alternations ON SP SET min OFF Hy 1 2 max GIR GIA ALARM Scale Analog CAL SET 1 2 3 4 I...

Page 2: ...ge 7 Max value reset page 7 Display of min value page 7 Min value reset page 7 Display setting of making point switching point for switching output 1 or 2 page 7 Display setting of breaking point hyst...

Page 3: ...l supply the GIR1002 FR factory tested and completely calibrated Before you can actually use it make sure to configurate your GIR1002 FR for your special application Please also refer to the chapter C...

Page 4: ...connected to a suitable RC element for example RC220 Electric connection and commissioning of the device must be carried out by trained and skilled personnel Wrong connection may lead to the destructi...

Page 5: ...min selectable Interface RS485 Power supply 230VAC or as specified on device housing Power consumption 5W 230VAC 3W 24VDC 1 5W 12VDC Relay switching power 10A 250V AC ohmic load use RC element when sw...

Page 6: ...setting of making point button 1 4 1 2 3 7 display setting of switching point button 1 4 4 5 6 8 display setting of breaking point button 1 4 1 2 3 7 display setting of hysteresis button 1 4 4 5 6 8 S...

Page 7: ...in case of alarm LED min LEDflashing Min value reset To be called up by pressing button 3 for approx 4s Display please refer to display of min value Display setting of makingpoint switching point for...

Page 8: ...briefly no more than 1 s When pressing these buttons for a longer time over 1 s the values will start scrolling upwards respectively downwards After scrolling of approx 150 digits the scrolling speed...

Page 9: ...int to the left or button 3 shift decimal point to the right to set desired decimal point position acknowledge position by pressing button 4 If decimal point is positioned to the right of the lowest d...

Page 10: ...urrent of 20mA LED max will start flashing acknowledge by pressing button 4 8 LEDs Analog min and LED CAL are illuminated The display will show the measuring value stored in the GIR1002 FR for which a...

Page 11: ...rmissibe range This fault code indicates that the min display value of 1999 digits has fallen below Remedies FE 4 will be reset as soon as the values on the display are back within their permissible r...

Page 12: ...This method as well as the fact that the interface operates in the half duplex mode ensure that up to 16 devices can be called via only one 2 wire line Interface configuration Type RS485 Transfer mode...

Page 13: ...s specifying the parameter and or value to be processed For assignments please refer to table 1 of chapter functional code ASCII Code 2FH Indicates the end of a data transfer process Each data transfe...

Page 14: ...g point Switching point 0 9 30H 39H 1 Switching output 2 Breaking point Hysteresis 0 A 30H 41H 1 Max alarm limit 0 B 30H 42H 1 Min alarm limit 0 C 30H 43H 1 Input signal 1 5 31H 35H 3 Position of deci...

Page 15: ...l The assignment is as follows 0 2Vss 1 300mVss 2 30mVss 3 TTL Data format 4 This format is used to call up and or program the position of the decimal point The assignment is as follows 0 no decimal p...

Reviews: