E35.0.01.6C-03 Manual for connection and operation of the EB 3000 page 12 of 19
Er.11: Value could not be calculated
Indicates a measuring value, needed for calculation of the display value, is faulty resp. out of
range.
For possible causes and remedies please read the operation manual of the
EASY
BUS
-sensor mod-
ule
Er.12: Received datasets invalid
Indicates, that the received
EASY
BUS
-sensor module datasets for this
EASY
BUS
-measurement
channel is invalid.
Possible causes:
-
EASY
BUS
-measurement channel with an error indication
Remedies:
- check
EASY
BUS
-sensor module
E.-23:
Transmission error: EASYBus-modul not responding
Indicates, that the attached
EASY
BUS
-sensor module does not responding an enquiry.
Possible causes:
-
EASY
BUS
-sensor module not connected or defective
- incorrect system initialisation or configuration
- uncouple / access of the sensor module via PC-software
(e.g. searching of GSOFT40K)
Remedies:
- check
EASY
BUS
-sensor module and connection
- perform new configuration of the EB3000
- check bus access via PC-software
E.-25:
Transmission error: CRC-Code wrong
Indicates, that the enquiry of an attached
EASY
BUS
-measurement channel has generated an CRC-
error.
Possible causes:
- multiple attached
EASY
BUS
-measurement channels have the same address
-
EASY
BUS
-sensor module defective
Remedies:
- perform new
EASY
BUS
-system initialisation
E.-38:
No acknowledge received (unknown operation call)
Indicates, that the selected function of the attached
EASY
BUS
-measurement channel (at present) is not
supported.
Possible causes:
- a working Logger accepts some settings only in stop-modus
- module = GIA20EB: alarm function not activated
Remedies:
- stop Logger
- activate alarm function of the GIA20EB beneath “out”
E.-41:
Data received within locked range
Indicates, that received data are not in a valid
EASY
BUS
-value range.
Possible causes:
-
EASY
BUS
-sensor module incorrect configured or defective
Remedies:
- check
EASY
BUS
-sensor module
E.-44:
Transmitted data modified removed
Indicates, that transmitted data varied from data, saved and acknowledged from
EASY
BUS
-sensor
module.
Possible causes:
- The transmitted data are outside of a permissible value range of the
EASY
BUS
-measurement channel.
- The
EASY
BUS
-sensor module has accepted a possible min / max value
and has send back this value to the EB3000.
Remedies:
---