page 11
2.1 MS WINDOW ( Measuring window )
B 600 M
B 600 - MS WINDOW
The measuring window has a factory setting
value of 100 ° .
This means, the B 600 can display an amplitude
value higher than 100 °.
You may change this setting, if you have
problems with the amplitude of certain watches
B 600 M
RUN MODUS : N FIXWERT
MST( STEP ) : 4 sec
LIFT ANGLE : 52 °
MST (STEP) : 21600 / h
MS WINDOW : 100 °
A B CCC
If the escapment noise is not clear
(because of mechanical problems of the watch,
the noise signal may look as in figure 5 .
problems with the amplitude of certain watches.
SELECT ACTUAL RUN PARAMETERS
figure 5
The value ( t ) is not always identically.
The B 600 may display a rong Amplitude ,
e.g 170 ° instead of 250°.
In this case you have the possibility to change
the value called ( MS WINDOW ).
t
figure 5
The value ( MS WINDOW ) is the minimum amplitude
the B 600 can display.
If you increase the value to 220°, a real lower value
will not be displayed. The place where amplitude
normally is displayed, remains empty.
t = unstable
The MS WINDOW cuts a part of the escapment
noise signal.
100°
150°
180°
200°
220°
escapment
B 600 M
RUN MODUS
AUTMATCL
100
MS WINDOW
escapment
signal
fi
6
SELECT ACTUAL RUN PARAMETERS
RUN MODUS : AUTMATCL
MST( STEP ) : 4 sec
LIFT ANGLE : 52 °
N ( STEP ) : 21600 / h
MS WINDOW : 180 °
Operating - Manual B 600 M / Q
figure 6
SELECT ACTUAL RUN PARAMETERS
To save the new value of MS WINDOW when in the above menu,
Press ( PRINT ) first and then ( MENU - )