background image

page  7

1.5   Set Time & Date

a. Execute step 1.2
b Press

NEXT

Programmed Time and Date is displayed

B 600 - SYSTEM CONFIGURATION

b. Press   - NEXT   

Programmed Time and Date is displayed.
Cursor stays on hour.

c. Press   - MENU + or MENU -

Until the desired hour appears.

d. Press   - NEXT   

Cursor stays on minute.

e.  Press   - MENU + or MENU -

Until the desired minute appears.

f. Press   - NEXT   

Cursor stays on second.

g.  Press   - MENU + or MENU -

Until the desired second appears.

h.  Press   - NEXT   

Cursor stays on day.

y

y

i.  Press   - MENU + or MENU -

Until the desired day appears.

k.  Press   - NEXT   

Cursor stays on month.

l.  Press   - MENU + or MENU -

Until the desired month appears.

m. Press   - NEXT   

Cursor stays on year.

n. Press   - MENU + or MENU -

Until the desired year appears.

o. Press   - PRINT /  PRINT /  (MENU -)

The selection will be stored.

1.6   Set the desired test parameters 

a. Switch on the B 600 and wait until 

the screen  1  is displayed :

B 600  M

WATCHTYPE :

PENDULUM

MECHANIC

1 7 S t th d

i

d t

t

t

1

SELECT  ACTUAL  RUN  PARAMETERS

MECHANIC

1.7   Set the desired test parameters 

for mechanical watches.

a. Press (Menu + ) to select the watchtype:

MECHANIC

B 600  M 

b. Press  (NEXT) screen  2  is displayed

By pressing (MENU + ) you may select

AUTOMATCL  beat selection or N FIXED

SELECT ACTUAL RUN PARAMETERS

RUN MODUS : 

AUTMATCL

N  FIXED       

2

AUTMATCL

The B 600 detects 
automatically  the beat 
of the watch

SELECT  ACTUAL  RUN  PARAMETERS

of the watch.

N  FIXED       

The beat can be set  
manually from :
3600  to  36000 by steps.

1 8 Set the desired MST

B 600  M

RUN MODUS :   N FIXED
MST( STEP )  :          4 sec
LIFT ANGLE   :        50 °

3

Operating - Manual B 600 M / Q

1.8   Set the desired MST

LIFT ANGLE and 
BEAT  FREQUENCY

See next page for explanations !

SELECT  ACTUAL  RUN  PARAMETERS

N ( STEP )      :                   / h  

21600

Summary of Contents for B 600 M

Page 1: ...ervibrograf ch www greinervibrograf ch Vibrograf B 600 M Q Vibrograf B 600 M Q User manual User manual User manual User manual Operating Manual B 600 M Q Helmut Klein GmbH Fritz Neuert Strasse 31 7518...

Page 2: ...e 4 page 5 page 6 page 7 EXPLANATION OF TEST PARAMETERS DISPLAY OF WATCH SIGNAL AUTO CALCULATION OF AMPLITUDE MEASURING WINDOW page 7 page 8 page 9 page 10 TESTING CLOCKS TESTING QUARTZ WATCHES TESTIN...

Page 3: ...NFIXED 21600 MT 2 LA 52 26 s d 218 0 2ms On Off switch and volume adjust for internal speaker out of beat value Amplitude value time deviation sign T l b l Rear view Speaker parallel interface female...

Page 4: ...or printout 2 Switches to the double beat 1 Starts the printout 2 Switches to the single Goas to the next MENU point On Off switch and volume adjust for internal speaker Di l 2 Switches to the single...

Page 5: ...tts the watch sound noise to the cristal inside the microphone The cristal converts the noise in electrical signals Through the built in electronic and the cable the signals reaches the B 600 The sign...

Page 6: ...LANGUAGE NEXT SET TIME DATE BACK DEFAULT ALL PARAMS MENU LOAD SAVED PARAMS MENU SAVE ACTUAL PARAMS PRINT RETURN W O CHANGES WAITING FOR YOUR SELECTION 1 3 Change the language a Execute step 1 2 b Pres...

Page 7: ...NU The selection will be stored 1 6 Set the desired test parameters a Switch on the B 600 and wait until the screen 1 is displayed B 600 M WATCHTYPE PENDULUM MECHANIC 1 7 S t th d i d t t t 1 SELECT A...

Page 8: ...4 sec LIFT ANGLE MS WINDOW 100 The value of lift angle depends on the construction of the watch It is used in the B 600 to calculate the amplitude Most watches have a lift angle of 50 52 It can be se...

Page 9: ...PE PENDULUM MECHANIC In this menu press FREEZE press START STOP to return to the main menu DIAGRAM DOUBLE BEAT PRESS START STOP KEY TO COMPLETE For more ditails DIAGRAM SINGLE BEAT press PRINT to disp...

Page 10: ...plitude 230 t 0 009 s Amplitude 230 Remark The value of the time t depends on the mechanical datas of escapment from the watch This value is always measured by the B 600 and the amplitude is calculate...

Page 11: ...t always identically The B 600 may display a rong Amplitude e g 170 instead of 250 In this case you have the possibility to change the value called MS WINDOW t figure 5 The value MS WINDOW is the mini...

Page 12: ...place the light barrier max 5 mm from the arm pendulum Light barrier the arm pendulum b Switch the B 600 to PENDULUM pressing MENU B 600 M WATCHTYPE max 5 mm SELECT ACTUAL RUN PARAMETERS MECHANIC PEN...

Page 13: ...ches The shown right side is equipped with 3 capture systems red point capacitif LCD quartz 4 19MHz metal pin microphone for quartz 32 KHz yellow point inductif capture of motor impulses yellow point...

Page 14: ...watch on the yellow point Each motor impulse will be displayed with A arrow in the lower part of the display For example quartz watches with 1 second impulses Or quartz watches with inhibition of 10 s...

Page 15: ...2 5 10 20 50 100 200 500 Same watch 0 12 s d The higher you set the DSP Y GAIN value The higher is the resolution of the displayed grafic The higher is the resolution of the displayed grafic B 600 Q 2...

Page 16: ...impression that the watch signal is too weak and the grafic gives problems How to adjust Turn the Knob Level in front of the B600 in the position one o clock point to point Lay a good working watch c...

Page 17: ...ot active paralell printer connector Display 1 LCD Display adjustment of brightness with potentiometer on the rear Functional sequence 2 LED s printing and measuring Measuring possibilities mecanical...

Reviews: