page 8
B 600 M
RUN MODUS : N FIXED
B 600 - EXPLANATION OF TEST PARAMETERS
Press ( Next )
to move down in this menu.
Press ( Back )
SELECT ACTUAL RUN PARAMETERS
MST( STEP ) : 4 sec
LIFT ANGLE : 50 °
N ( STEP ) : / h
MS WINDOW : 100 °
21600
(
)
to move up in this menu.
Press ( Menu + ) or ( Menu - )
to change a value in the marked line.
Press and hold ( MENU + ) to encrease
Press and hold ( MENU - ) to decrease
the value continously
the value continously
B 600 M
RUN MODUS : AUTMATCL
MST( STEP ) : sec
MST( FINE )
4
Measuring time : ( MST )
The result of time deviation will be
desplayed after the set value of ( MST )
and will be the meanvalue of the ( MST )
l
i MST (STEP )
SELECT ACTUAL RUN PARAMETERS
LIFT ANGLE : 50 °
MS WINDOW : 100 °
values in MST (STEP ) are:
4, 8, 12, 30, 60, sec.
values in MST ( FINE ) are:
2, up to 960 sec. in steps of 2
Lift angle: (50 - 52°)
50
B 600 M
50
RUN MODUS : AUTMATCL
MST( FINE ) : 4 sec
LIFT ANGLE : °
MS WINDOW :
100 °
The value of lift angle depends on
the construction of the watch. It is used
in the B 600 to calculate the amplitude.
Most watches have a lift angle of 50°-52°.
It can be set from 30° up to 70°.
If the value of lift angle is set lower than
the value of the measured watch, the
result of amplitude will be lower than
SELECT ACTUAL RUN PARAMETERS
MS WINDOW : 100
result of amplitude will be lower than
it should be. If it is set higher the result of
amplitude will be higher.
Beat frequency N (STEP) or N (FINE)
In RUN MODUS :
AUTMATCL
21600
50
B 600 M
RUN MODUS : AUTMATCL
MST( STEP ) : 4 sec
LIFT ANGLE : 50 °
MS WINDOW: 100°
In RUN MODUS :
AUTMATCL
the beat frequency is automatically
detected and set from the B 600 to a
standard beat. Standard beats are :
3600, 4800,
7200,
7920,
8100,
8640.
9360,
9440,
9720,
9760,
14400, 16200, 18000, 19800, 21600,
28800
36000
Operating - Manual B 600 M / Q
SELECT ACTUAL RUN PARAMETERS
28800, 36000,
In RUN MODUS
( N FIXED)
The beat can be set manually from
3240 up to 39600 in N ( FINE )