15
is recorded.
6 - CHANGING SET VALUES:
When the temperature set values are changed, the ball set or bottom set key is pressed first.
The set value is indicated on the display as S-1 ..... S-6. The change is made with the arrow
keys. Use the top set or bottom set key to move to the other set value. The upper group values
are entered in order of S-1 / S-2 / S-3. The lower group temperatures are entered as S-4 / S-5
/ S-6.
The time setting operation is done with only the arrow keys. There is a waiting time of 3 seconds
for recording.
7 - STEAMING:
Steam can not be given until the oven reaches the parameter temperature after the system has
started. When the steam input is ready, the user will be given steam by pressing each key.
8 - LAMP CONTROL:
The lamp relay can be turned on and off by pressing the lamp button. If the lamp relay
parameters are set, it can also be operated on time.
9 - ERROR INDICATORS:
The error indicators for this device are described below.
Mean: The box is active if the temperature sensor can not read.
Tc1: 1. Termocouple is faulty. The system is automatically shut down.
Tc2: 2nd thermocouple failure. The system is automatically shut down.
Tc3: 3rd thermocouple failure. The system is automatically shut down.
Tc4: 4. Termocouple is faulty. The system is automatically shut down.
Tc5: 5. Termocouple failure. The system is automatically shut down.
Tc6: 6. Termocouple failure. The system is automatically shut down.
10 - AUTOMATIC OPENING:
When my site is off, press the start time key. The time on the set indicator starts to show the
automatic turn-on time which is not displayed until then. If you want to change the time, press
the Start key until the setting is written on the screen. Then the setting is made with the arrow
keys and saved with the Start Time key.
When the time is equal to the actual time displayed on the counter display, the device
automatically opens and starts running.
To cancel When the time set display has time to open, it is necessary to press the time stop key
once. This will turn off the Duration set indicator and cancel the automatic power-on.
11 - ELECTRICAL CONNECTIONS:
The connection pattern must be observed when the device is electrically connected. The device
supply voltage is 220 VAC 50 Hz.
The device relays should not be used to drive loads above 1 amp. The external contactor must
be used for this operation.
The device can only measure the temperature with a J-type thermocouple. Measurement with
other types of sensors will be incorrect.
Electrical connections should be made only by technical personnel. The system energy must be
disconnected before the exchange operations to be made in connection.
The system should not come into direct contact with water.