15
93-97 / 00220
DC Current
Range
Resolution
Accuracy
Voltage Burden
300
µ
A
0.1
µ
A
±
(1% rdg. + 2 digits)
200mV max.
3mA
1
µ
A
±
(1.2% rdg. + 2 digits)
2V max.
30mA
10
µ
A
±
(1% rdg. + 2 digits)
200mV max.
300mA
0.1mA
±
(1.2% rdg. + 2 digits)
2V max.
10A
10mA
±
(2% rdg. + 3 digits)
2V max.
Overload Protection: 1A/500V fast blow fuse for
µ
AmA input, 10A/500V
for A input
AC Current
300
µ
A
0.1
µ
A
200mV max.
3mA
1
µ
A
±
(1.5% rdg. + 3 digits)
2V max.
30mA
10
µ
A
200mV max.
300mA
0.1mA
±
(2% rdg. + 3 digits)
2V max.
10A
10mA
±
(2.5% rdg. + 5 digits)
2V max.
Overload Protection: 1A/500V fast blow fuse for
µ
AmA input, 10A/500V
for A input
Frequency Response: 40Hz~500Hz