2-12
CHAPTER 2 Checks and Preparation
(2) Attach the expansion terminal base to the GLT400.
* Pay attention to the protrusions when direct-connecting.
● Please direct-connect the extension terminal bases carefully so as not to bend the protrusions next to
the connector.
● Please handle the GLT400 in a horizontal state until the extension terminal bases are fixed with the
connecting plate.
(3) Install it with the supplied screws.
Hold the expansion terminal base at a height of 33 to 34 mm, align the height of the expansion terminal
base with the height of the GLT400, and tighten the screws.
* Recommended tightening torque: 14 kgf/cm
M4×6 flat screw
Prepare the spacer with a height of 33 to 34 mm, and align the height of
the expansion terminal base with the height of the GLT400.
Expansion terminal base
GLT400
Spacer
If the height of the GLT400 and the expansion terminal base are not aligned, the direct-connected
connector may be damaged due to the load applied to the connector.