CHAPTER 4 Settings and Measurement
(2)-5 EU (Scaling settings)
This is used to convert the measured signals to other units.
(1)
<For voltage input>
(5)
(2)(3)(4)
(6)
<For temperature input>
(2)(3)
(6)
(7)
Setting
Description
(1) EU Function
Sets the scaling function to ON or OFF.
(2) Meas. Value (Upper/Lower)
Sets the upper and lower limits of values to be converted.
* For temperature input, there is no distinction between upper and lower limits. See the setting
examples shown below for details.
(3) EU Output Value (Upper/
Lower)
Sets the upper/lower limit output values after conversion.
* For temperature input, there is no distinction between upper and lower limits. See the setting
examples shown below for details.
(4) Dec pt
Sets the decimal point position for an EU output value.
(5) Select
Selects a specific engineering unit classification. (The following are available.)
Cur, Length, Area, Volume, Velocity, Accel, Freq, Mass, Energy, Pressure, Flow , Temp, Strain,
Brightness,Concentration
(6) Unit
Selects a unit to be used after conversion. A unit displayed in this field belongs to the
classification selected in "Select." To set a unit not displayed in this field, set arbitrary text in
"Arbitrary unit." The setting selected in this field is displayed in "Arbitrary unit."
(7) Reads the current temperature
measurement value
Substitutes the current measurement value into (2) Measurement value and (3) EU output value.
* The value is not substituted when burnout occurs or the scale is exceeded.
● If a message window opens, follow the instruction in the message to change the setting value.
● The Scaling function performs calculation using a ratio of the Meas. Value and EU Output Value settings.
● The digital display shows “++++/----” when the converted value cannot be processed by the GLT400.
● The span may be changed depending on the Scaling settings.
● For temperature input, the offset setting for an input value is used.