8
User Manual
Function Inspection
1. Press the
Default
button
on the front panel
to restore the instrument to its default
configuration.
2. Connect the ground alligator clip of the probe to the “Ground Terminal” under the
probe compensation signal output terminal.
Compensation Signal Output Terminal
Ground Terminal
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
“Compensation Signal Output Terminal” of the probe.
4. Press the
Auto Setup
softkey.
5. Observe the waveform on the display. In normal condition, the display should be a
square waveform as shown in the figure below:
Figure 5 Function Inspection
6. Use the same method to test the other channels. If the square waveforms actually
shown do not match that in the figure above, please perform
“Probe Compensation”
in the next section.
WARNING
To avoid electric shock during the use of probe, please make sure that the
insulated wire of the probe is in good condition and do not touch the metallic
part of the probe when the probe is connected to high voltage source