User Manual
161
To Enable Quick-Cal
Quick-Cal can calibrate the measure deviation which cause by the temperature. If the
temperature that the oscilloscope is working is not stable, do the following steps:
1. Press the
Utility
button on the front panel to enter the UTILITY function menu.
2. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
3. Press the
Quick-Cal
to select
On
.
If the Quick-Cal has been set to
On
before powering on, and the current vertical scale is
2mV or 5mV
of any channel, the oscilloscope will perform the quick calibration procedure
when it powers on. During the calibration, most of the buttons are disabled.
Also, when the Quick-Cal is
On
and the
current vertical scale is 2mV or 5mV
of any
channel, the oscilloscope will perform the quick calibration procedure if the operating
temperature changes by more than 2 degree.