To Do Self Calibration
The self-calibration program can quickly make the oscilloscope reach the best working
state to get the most precise measurement values. You can perform self-calibration at any
time especially when the change of the environment temperature is up to or more than
5
℃
. Make sure that the oscilloscope has been warmed up or operated for more than 30
minutes before the self-calibration.
Do the following steps to do self calibration:
1. Disconnect all the input channels.
2. Press the
Utility
button on the front panel
,
and then press the
Do Self Cal
softkey,
and the oscilloscope will pop-out the message box shown as below:
Figure 93 Do Self Cal
3. Press the
Single
button on the front panel to perform the self calibration program.
During the calibration, most of the keys are disabled.
4. When the self calibration program is finished, it will pop-out the message “
press
Run/Stop key to exi
t”. Press the
Run/Stop
button on the front panel to exit the
calibration interface.