GMC-I Messtechnik GmbH
6.3.3
Ending the Measurement and Discharging
➭
End the respective measurement by pressing the
Stop
softkey.
After ending the measurement, any remaining re-
sidual voltage (Uext) is displayed (message:
„Discharge ...“), which may result from cable ca-
pacitance. The instrument’s internal 1 M
resis-
tor causes rapid discharging.
Contact to the motor coils must be retained to this end.
Do not disconnect the object until voltage has fallen
below 25 V, i. e. message „Discharge ...“
disappears.
Note
Voltage measurement in the
RISO
or
COIL
switch position
is used primarily for interference voltage detection before
the respective measurement.
Use the V , V
or V switch position in order to per-
form precise voltage measurements.
+
–
0 V!
Discharge