52
GMC-I Gossen-Metrawatt GmbH
triG = out
Measurement values are stored if at least one measure-
ment value occurs which lies within the limits for H-triG and
L-triG, and if one of the two limit values is violated thereaf-
ter.
SEt
↵
V
triG
↵
V
↵
VU
out
↵
H-triG
...
triG = in
Measurement values are stored to memory if the following
conditions are met: at least one measurement value occurs
which lies outside of the limits for H-triG or L-triG, and if
one of the two limit values is again violated thereafter.
SEt
↵
V
triG
↵
V
↵
VU
in
↵
H-triG
...
triG = Sto
–
ou
Only those measurement values are stored to memory,
which do not lie within the limits for H-triG and L-triG.
SEt
↵
V
triG
↵
V
↵
VU
Sto
–
ou
↵
H-triG
...
triG = Sto
–
in
Only those measurement values are stored to memory,
which do lie within the limits for H-triG and L-triG.
SEt
↵
V
triG
↵
V
↵
VU
Sto
–
in
↵
H-triG
...
22.2.1 Trigger Function Parameters
Upper Limit
H-triG
, Lower Limit
L-triG
Each trigger function is assigned its own upper and lower
limits as trigger parameters. The upper limit should be
greater than the respective lower limit.
The predetermined trigger threshold is evaluated according
to the selected measuring range, independent of whether
or not automatic measuring range selection is activated.
Entering upper and lower trigger thresholds in digits:
SEt
↵
V
triG
↵
V
↵
in
H-triG
↵
XXXXXX
VU
↵
L-triG XXXXXX
VU
↵
.
Note!
☞
Values for H-triG and L-triG are also used as trigger
parameters for cont in, events or mains. The table in
chapter 17.1 contains examples for the entry of trig-
ger thresholds in digits.
Pre-Trigger:
PrEtr
If this function has been activated (on), measurement val-
ues are stored to memory immediately after the memory
mode has been activated dependent upon the time trigger
(see below). If this function has been deactivated (OFF),
only those measurement values which exceed the trigger
threshold are stored to memory when the memory mode is
activated.
SEt
↵
V
triG
↵
V
MAinS
↵
rAnGE: XXXX V
↵
H-triG: XXXXXX
↵