DT Pump Head User’s Guide 

 

Figure 5. Fix the Cartridge 

For multiple channel application, Please make sure the tubing length in 

the cartridge and the stress on the tubing are almost  the same, to 

minimize the flow rate difference between the channels. 

 

Note: 

1)  For multiple channels, the flow rate will not be identical. The 

difference might be minimized by fine adjusting the Set Screw for 

occlusion adjustment. 

2)  When the pump is running, the rollers squeeze the tubing to 

deliver the liquid. Please check the tubing regularly and replace it 

before it wears out. 

3)  Because of the tubing wear, the flow rate would become smaller 

after a period of use, adjust the speed or recalibrate the pump 

when necessary. 

4)  Do not tighten the Set Screw too much. As long as the pump head 

can deliver liquid, the screw is tight enough. 

Summary of Contents for DT10-18

Page 1: ...GOLANDER PUMP DT Pump Head User s Guide info golanderpump com http golanderpump com 1 678 587 8806...

Page 2: ...t designed for nor intended for use in patient connected applications including but not limited to medical and dental use Note Only use matched tubing with the pump head to ensure optimum performance...

Page 3: ...channels PPS material case and stainless steel rollers More compact compared to multiple channel YZ or YT pump heads 3 Specifications DT15 Rollers 4 Roller Material Stainless Steel Cartridge Material...

Page 4: ...8 1 4 5 6 7 3 2 Figure 1 DT15 Pump Head and DT10 Pump Head 1 Front panel 2 Cartridge for tubing 3 Set screw for occlusion adjustment 4 Back panel 5 Support metal rod 6 Hole for Allen key 7 Snap fit H...

Page 5: ...the pump head insert the shaft tang of the pump head into the slot in the shaft on the drive Then align the pump onto the two alignment pins on the drive 1 Remove the cartridge s press the pump head...

Page 6: ...the desired distance between the two stops on the tubing to around 105mm for optimum performance as shown on Figure 3 105mm Figure 3 Stops for DT10 Pump Head 4 Load the tubing into the slot of the ca...

Page 7: ...identical The difference might be minimized by fine adjusting the Set Screw for occlusion adjustment 2 When the pump is running the rollers squeeze the tubing to deliver the liquid Please check the tu...

Page 8: ...T10 28 67 DT10 48 87 5 DT10 88 102 DT15 DT15 14 67 DT15 24 87 5 DT15 44 102 6 Warranty The product comes with one year labor and parts warranty The limited warranty does not cover any damage that is c...

Page 9: ...0 680 1 600 1 960 3 240 5 380 40 0 092 0 364 1 360 3 200 3 920 6 480 10 760 60 0 138 0 546 2 040 4 800 5 880 9 720 16 140 80 0 184 0 728 2 720 6 400 7 840 12 960 21 520 100 0 230 0 910 3 400 8 000 9...