DT Pump Head User’s Guide 

pressure. The tubing length is 2 meters, 1m for inlet and 1 for outlet. 

Table 1. DT15 Flow Rate (mL/min) 

Speed 

(rpm) 

Tubing Size 

19# 

16# 

25# 

17# 

100 

47 

78 

177 

267 

300 

140 

235 

530 

800 

600 

280 

470 

1060 

1600 

 

Table 2. DT10 Flow Rate (mL/min) 

Speed 

(rpm) 

Tubing Size (mm) 

0.13 

0.25 

0.5 

0.76 

1.22 

1.52 

0.1 

0.000 

0.001 

0.003 

0.008 

0.010 

0.016 

0.027 

20 

0.046 

0.182 

0.680 

1.600 

1.960 

3.240 

5.380 

40 

0.092 

0.364 

1.360 

3.200 

3.920 

6.480  10.760 

60 

0.138 

0.546 

2.040 

4.800 

5.880 

9.720  16.140 

80 

0.184 

0.728 

2.720 

6.400 

7.840  12.960  21.520 

100 

0.230 

0.910 

3.400 

8.000 

9.800  16.200  26.900 

Speed 

(rpm) 

Tubing Size (mm) 

1.65 

2.4 

2.79 

3.17 

13# 

14# 

0.1 

0.028 

0.036 

0.041 

0.057 

0.082 

0.005 

0.018 

20 

5.660 

7.200 

8.200  11.400  16.400 

1.000 

3.600 

40 

11.320  14.400  16.400  22.800  32.800 

2.000 

7.200 

60 

16.980  21.600  24.600  34.200  49.200 

3.000  10.800 

80 

22.640  28.800  32.800  45.600  65.600 

4.000  14.400 

100 

28.300  36.000  41.000  57.000  82.000 

5.000  18.000 

 

 

Summary of Contents for DT10-18

Page 1: ...GOLANDER PUMP DT Pump Head User s Guide info golanderpump com http golanderpump com 1 678 587 8806...

Page 2: ...t designed for nor intended for use in patient connected applications including but not limited to medical and dental use Note Only use matched tubing with the pump head to ensure optimum performance...

Page 3: ...channels PPS material case and stainless steel rollers More compact compared to multiple channel YZ or YT pump heads 3 Specifications DT15 Rollers 4 Roller Material Stainless Steel Cartridge Material...

Page 4: ...8 1 4 5 6 7 3 2 Figure 1 DT15 Pump Head and DT10 Pump Head 1 Front panel 2 Cartridge for tubing 3 Set screw for occlusion adjustment 4 Back panel 5 Support metal rod 6 Hole for Allen key 7 Snap fit H...

Page 5: ...the pump head insert the shaft tang of the pump head into the slot in the shaft on the drive Then align the pump onto the two alignment pins on the drive 1 Remove the cartridge s press the pump head...

Page 6: ...the desired distance between the two stops on the tubing to around 105mm for optimum performance as shown on Figure 3 105mm Figure 3 Stops for DT10 Pump Head 4 Load the tubing into the slot of the ca...

Page 7: ...identical The difference might be minimized by fine adjusting the Set Screw for occlusion adjustment 2 When the pump is running the rollers squeeze the tubing to deliver the liquid Please check the tu...

Page 8: ...T10 28 67 DT10 48 87 5 DT10 88 102 DT15 DT15 14 67 DT15 24 87 5 DT15 44 102 6 Warranty The product comes with one year labor and parts warranty The limited warranty does not cover any damage that is c...

Page 9: ...0 680 1 600 1 960 3 240 5 380 40 0 092 0 364 1 360 3 200 3 920 6 480 10 760 60 0 138 0 546 2 040 4 800 5 880 9 720 16 140 80 0 184 0 728 2 720 6 400 7 840 12 960 21 520 100 0 230 0 910 3 400 8 000 9...

Reviews: