26
TROUBLESHOOTING
Delivery Unit
Problem
Probable cause
Solution
- Compressor disconnected.
- Inlet pressure below specified
(80 PSI / 5,52 BAR).
- Bio-system reservoir run out of
water.
- Chair fuse burned.
- Main or chair switch is off.
- Handpiece is not working
- Handpiece with low speed
- When Bio-system is
operated no disinfectant come
from handpiece terminals.
- Plug the compressor in.
- Adjust inlet pressure (80 PSI
/ 5,52 BAR).
- Put disinfectant in the
reservoir.
- Replace chair fuse.
- Switch main/ chair switch on.
- Reservoir run out of water.
- Compressor disconnected.
- No water from syringe.
- Put filtered water in reservoir.
- Plug compressor in.
- Compressor disconnected.
- No air from syringe
- Plug compressor in.
- Chair’s fuse burned
- Main or chair switch is off.
- Replace chair’s fuse
- Switch main/ chair switch on.
- Lack of bicarbonate in
rerservoir.
- Spout or reservoir outlet is
clogged.
- Bicarbonate excess in
reservoir.
- Wrong jet position.
Bicarbonate jet
(GLXF FULL, GLXC FULL, GLXF, GLXC, GLF
JET and GLC JET Delivery Units)
Not enough bicarbonate in
jet.
- Add bicarbonate to reservoir
(max. 40g).
- Clean clogged parts
- Remove excess.
- Change position
- Compressor disconnected
-
Low pressure in jet.
- Plug compressor in
- Water valve is closed.
- Jet water reservoir is empty.
-
Few water in jet.
- Open valve.
- Put water in reservoir.
- X ray viewdoes not
work