MAINTENANCE
How to provision the reservoirs
Water - Syringe / Handpieces
Remove the reservoir (45) uncoiling it on
clockwise and make the replacement of water. After
the replacement put it back coiling on anticlockwise.
Always use filtered water or aseptic products.
Bio-System
Remove the reservoir (44) uncoiling it on
clockwise and make the replacement. Use a
chlorinated water solution 1:500
Preparing the solution:
From a solution of hypochlorite of sodium at 1%
(solution of Milton, Virex), a solution of chlorine at
500 p.p.m. is prepared.
How to prepare the solution:
Take 25ml of
hypochlorite of sodium at 1% and dilute it in 500 ml
of water (1 to 20). Such solution should be prepared
daily.
IMPORTANT:
Follow this proportion strictly to avoid damages in the equipment and to have an efficient result in the
disinfection.
Water - Scaler / Bicarbonate Jet
(Syncrus GLXF FULL, GLXC FULL, GLXF, GLXC, GLF SONIC, GLC SONIC,
GLF JET and GLC JET Delivery Units)
Remove reservoir (46) unscrewing it anticlockwise and replace water or disinfectant liquids. After
replacement place it again screwing it clockwise.
45
43
44
Level adjustment of the Delivery
Unit´s box
(Syncrus GLXF FULL, GLXF, GLF SONIC, GLF
JET, GLF OPTI and GLF Delivery Units)
Loosen the screws (45), remove the covers (46)
and remove the superior cover (47), soon after make
the level adjustment of the box delivery unit through
the screw (48). Make the adjustment using a level,
being taken as reference the arm (49).