OPERATION
Optilight LD
(Syncrus GLF OPTI and GLC OPTI Delivery Unit)
To initiate a activated cycle, press the timer
trigger (40), which generates a short beep every 10
seconds and a long beep at the end of a 90 second
cycle.
• To interrupt a activated cycle just activate the
timer trigger again (40).
IMPORTANT:
• Keep the light conductor tip (42) at least 2mm
away from the restoration.
• Keep the light conductor (42) always protected
by an expendable PVC film, which must be changed
for every patient. This procedure protects the light
conductor from scratches and other residues.
• Use the polymerization time recommended by the compound resin manufacturer and always perform
restorations in incremental layers with a maximum thickness of 2mm.
Instructions for equipment operation
WARNING
• Never aim the blue light beam towards the eyes
• Use the eyesight protection (41)
In order to protect the eyes, the eyesight protection (41) filters only the blue light used for the resins
polimerization, and it allows ambient light to pass through.
Remove the upper lid (16), add 20 to 40 sodium
bicarbonate, this quantity is enough to meet
prophylaxis needs.
ATTENTION
In order to avoid clogging, don’t add more
than 40g of bicarbonate.
Bicarbonate level can be checked through the
transparent lid.
To adjust the bicarbonate jet power, use the
water adjustment valve (23); to increase pressure, turn it clockwise, to decrease, counterclockwise.
Equipment operation
Bicarbonate jet removes dark stains from teeth, caused by tobacco, coffee, tea, etc. related to plaque.
In order to obtain better results of its use, we recommend a distance of 5mm between the handpiece
and the teeth, with an angle from 30 to 45º, describing circular movements over the teeth.
The bicarbonate jet must be directed to the occlusal edge and not to the gingival edge in order to
avoid an unpleasant sensation to the patient.
Bicarbonate Jet
(Syncrus GLXF FULL, GLXC FULL, GLXF, GLXC, GLF JET and GLC JET Delivery Unit)
40
41
42
16