TECHNICAL SPECIFICATIONS
01 - Product’s identification label
02 - Attention: Hot surface
03 - Fuse 15A (F2-F3)
05 - Attention: This equipment leaves the manufacturer with the selecting key of voltage at 220V
being so be sure that the net is compatible before connecting. If the voltage will be 127V please
invert the voltage selection switch position.
06 - Fuse 08A (F4)
07 - Before first use, please remove the protector film, which wraps this tray.
08 - Fuse 0,5A (F1)
09 - Maximum power per resistance: 800W - Warning: 2 Resistances
10 - Before begin the cycle, provision with 200 ml of distilled water
Contents of the marks accessible and not accessible (Legends)
INSTALLATION
Preparing the autoclave (Steps previous to operation)
Installing or replacing fuses
NOTE:
With a Phillips screwdriver, unscrew the lid (09)
of the electronic board compartment, then remove
the broken fuse and place the new fuse (19) in the
position indicated.
ATTENTION:
Before installing or replacing fuses,
unplug the electric power cord.
NOTE: Fuses that come with the Autoclave:
- 2 15A fuses installed in the equipment (F2/F3) +
1 spare;
- 1 08A fuse installed in the equipment (F4) + 1 spare;
- 1 0.5A fuse installed in the equipment (F1) + 1 spare.
ATTENTION:
- After performing the first replacement, it is
advisable to always have spare fuses at hand.
- All fuses must be installed independently from
the grid voltage.
Voltage change
For your safety, the autoclave is factory-set to
220V. For this reason, you must check the suitability of
the electricity supply. If the electricity network available
is 127V, change the voltage switch.
Change the voltage switch (08) according to the
local electricity network (127 or 220 V).
WARNING:
Don’t change the voltage while the device is on.
Then, connect the equipment to the electricity
supply and proceed as follows.