5
WARRANTY OF THE EQUIPMENT
IMPORTANT SAFETY CONSIDERATIONS
- Never touch the steam external outlet (12) and, after the sterilization is finished, never touch
the internal autoclave’s surfaces. The inappropriate use may cause burns. GNATUS may not be hold
responsible for accidents caused by inadequate operation.
- Don’t allow patients, and most specially children, to touch the autoclave.
- Periodically check that the door’s door latch is correctly in place before operation. If this procedure
were not observed, the sealing ring may come off.
- Before using the autoclave, verify that the internal chamber is clean, also check if the filter (17A) is
completely clean, free of residues.
- When raising the door latch, the autoclave’s door must open without difficulty.
- Always make sure that the depressurization is complete: the manometer’s needle must be in the
“0” (zero) position. Never use force to open the autoclave!
- If one of the safety devices were activated (unexpected steam leak) usually caused by the
internal steam outlet blockage or by an obstruction in the pressure control valve, wait until the complete
depressurization before switching the autoclave off.
- Before turning your autoclave on, install the fuses included with the equipment.
If the voltage were changed, these fuses may be changed.
- Use only the hose supplied by GNATUS, or one similar, so as to prevent if from melting when in
contact with steam, which may cause the external steam outlet obstruction.
- We recommend you to read this manual entirely. Use is as consultation manual.
This warranty covers product’s manufacturing faults for the warranty period and standards specified
on the Warranty Certificate. The Warranty Certificate is attached to product, and must be completed by
the Authorized Technician when equipment is installed.
Doubts and information: GNATUS Call center (55-16) 2102-5000.
Maintenance of the equipment
The equipment doesn’t need preventive maintenance, however when there is any problem that
this manual can’t solve, ask for an authorized Gnatus technician.
List of pieces and circuit scheme
Gnatus Company declares that the supply of the circuit scheme, list of pieces or any other information
that propitiate technical attendance for the user, can be request if there is an agreement between the
user and Gnatus Company.
Protection against current overload
The equipment is protected from current overload by fuses.
GENERAL DATA