21
TROUBLESHOOTING
Imprevisto
Problem:
Probable Cause:
Solution:
- The autoclave can’t reach or
maintain the pressure high
- The voltage supplied is lower
than the one required by the
autoclave
- An electrician must check and
adapt the electricity supply
- Leakage in the safety
seal
- Disconnect the autoclave, and
remove the safety seal and the
anti-vacuum valve, clean them
and place them again. If it were
necessary, replace them.
- The door latch isn’t in position,
causing leakage
- Place the door latch pushing it
downwards
- Not enough water
-Check that the volume of water
is the specified
- Overloaded chamber
- Use up to 70% of the chamber,
leaving spaces between the
packages
- The autoclave doesn’t swi-
tch on
- The cord is disconnected
- Connect the cord to the so-
cket
- Power cut
- Wait until energy is back
- Bad contact in the socket
- The ON/OFF switch is OFF.
- Replace fuse(s)
- Burnt fuse(s)
- Replace socket
- Turn the ON/OFF button ON.
-The autoclave’s door can’t be
opened
-The autoclave isn’t yet depres-
surized
-Wait the autoclave depressu-
rization
ATTENTION:
Any other problem that could happen with its equipment, should only be repaired by an Authorized
Technician GNATUS.
The bad use, negligence or maintenance not made by technician authorized by the maker, will imply
in the loss of the warranty.
* In case of continous fails, request to the attendance section to the customer a
authorized technician Gnatus.