Buttons: Function:
= Low-resistance measurement with 200 mA DC test current
• Plug
= Low-resistance measurement via the shockproof socket and the red connection
• Fixed connection
= Low-resistance measurement via the mains connecting cable
(blue powerCON)/via the system installation and the red probe connection
• Probe - probe
= Low-resistance measurement between the red probe connection and
black probe connection
• Extension
= Low-resistance measurement via the shockproof socket and the non-heating
appliance plug C13 (TG Basic 2+)
• Extension
= Low-resistance measurement via the shockproof socket and adapter A2
(TG Basic 2)
• Rsl
calibration
= Calibration of measurement line resistances
/
= Insulation test with a test voltage of 250 or 500 V DC (max. 1 mA)
• LN – PE
= Isolationsprüfung zwischen Aktivleitern (gebrückt) und Schutzleiter
• LN – probe
= Insulation test between active conductors and the red probe connection
• Probe – PE
= Insulation test between the red probe connection and the protective earth
conductor contact on the shockproof plug test socket
• Probe – probe
= Insulation test between the red probe connection and black probe
connection
• L – N
= Insulation test between the L and N active conductors
WARNING! The test voltage of 500 V DC can destroy the test objects.
= Substitute leakage current measurement with an idle voltage of approx. 95 V AC (max. 2,5 mA)
• LN – PE
= Measurement of substitute leakage currents when an alternating current is
applied between the active conductor and the protective earth conductor.
• LN – probe
= Measurement of substitute leakage currents when an alternating current is
applied between the active conductor and the red probe connection.
• Probe – probe
= Measurement of substitute leakage currents when an alternating current
is applied between the red and black probe connections.
= Voltage measurement across the red and black probe connections. The mains voltage is
switched on at the shockproof test socket for this!
= Function test with
differential
,
touch current measurement
,
phase current
,
voltage
,
active power
and
standby power measurement
in one measurement step.
• Socket function test
= The mains voltage is switched on at the shockproof test socket.
Touch currents can be sampled at all conductive parts with the red sensor, while
differential current, voltage, phase current and active power measurements are conducted
simultaneously. The arrow keys can be used to activate additional functions, like standby
power measurement and polarity reversal of phase and neutral conductor.
• Additional functions in the TG Basic 2+
• Clip-on ammeter function test
= Differential current and phase current can be tested here with a
clip-on ammeter. The touch current can also be sampled with the red sensor.
• RCD triggering time function test with 30 mA