Connections (Figure 2)
Before connecting the test device to the mains voltage, the corresponding warnings in Section 1 ‘General
safety and warning notices’, the warnings on the mains adapter cable and, if applicable, the warnings on
the accessories must be heeded! The safety of the user, appliance and test object is only ensured when
they are used as intended! The protective earth conductor potential must be tested using the finger
contact after connecting the test device.
1. Mains connection Neutrik Power CON blue
2. Non-heating appliance plug connection for extension cables and RCD testing (TG Basic 2+), IEC 60320 C19
3. Connection for red measuring line „Probe“ or for connecting adapter A1 and A2
(Rpe measurement) (for TG Basic 2 only)
4. Connection for black measuring line „GND“
5. FAULT LED
6. GOOD LED
7. LCD-Display
8. Arrow keys
9. Buttons for individual measurements
and
10. On / Off button
11. Phase control lamp indicates mains voltage at the shockproof test socket
12. Finger contact LED
13. Finger contact
14. USB Typ C connection for communication (for TG Basic 2+only)
15. Connection for a clip-on ammeter (for TG Basic 2+ only)
16. Test socket
4.1
Mains connection, Neutrik powerCON blue (1)
Mains connection of the test device. The test device must only be operated on a 230 V AC +/-10 % 40 – 60 Hz,
grid fused with max. 16 A!
4.2
Connection, non-heating appliance plug TG Basic 2+ C13 (2)
To connect the included non-heating appliance line with an extension, or test non-heating appliance lines.
4.3
Connection, TG basic 2 (3)
To connect the included test cables/adapters (A1 or A2) with an extension or to test non-heating
appliance lines.
4.4
Connection, shockproof test socket (16)
To connect the test cable to a shockproof socket up to 16 A.
Do not connect protective earth conductor until immediately before the mains is switched on.
WARNING!!!
When connecting the test object to a test socket, a voltage dangerous to the touch can result on a
defective test object or touchable conductive parts not connected to the protective earth conductor.