9.10 Substitute leakage current measurement LN-PE
The „Substitute leakage current measurement“ is not available in the app!!!
Note!
The substitute leakage current measurement may only be carried out after the insulation resistance test
has been passed.
Note!
The substitute leakage current measurement is not a sufficient test if the appliance has switching elements
that switch all poles, as this measurement does not test the underlying current circuits. Also, any leakage
currents caused by inverters cannot be measured. As a result, active testing should generally be preferred
to this measurement.
Automatic test sequence in Omni-Remote: „Protection class I passive“.
For individual measurements on the appliance, select „LN - PE“ under
= Substitute leakage current
measurement.
Note: The measurement value is halved if you select „Symmetrical circuitry Yes“. This occurs if the test object
has Y capacitors from L and N to PE.