8. Trouble shooting: 

SYMPTOM 

PROBABLE  CAUSE 

REMEDY 

Does not turn 

on. 

1.  AC adapter is not securely 

connected 

2. Low battery 
3.  Indicator is damaged 

1.  Re-plug the AC adapter or rotate 

the plug to securely connect to 
the scale 

2.  Replace the batteries 
3.  Replace with a new indicator and 

perform calibration   

Ad

﹉﹉﹉

 

1.  The cable from platform to 

indicator is not correctly 
connected, or disconnected, or 
short circuit 

2.  Indicator is damaged 
3.  Load cell cable is broken 
4.  Load cell is damaged 

1.  Check the cable 
2.  Replace with a new indicator and 

perform calibration. 

3.  Return the scale for repair 

Ad

﹍﹍﹍

 

0

﹉﹉﹉

 

Indication is out of key zero range 

Reduce the weight on platform, till 
the indication is within the key zero 
range. 

0

﹍﹍﹍

 

Weight reading below Power On 
Zero limit. 

1.  Check whether an object is stuck 

between scale base, if yes, 
remove the object. 

2.  Perform zero calibration. 

﹉﹉﹉﹉

 

1.  Weight reading exceeds 

overload limit 

2.  The weight value cannot be 

displayed in the current unit of 
measure because it exceeds 6 
digits. 

1.  Reduce load on the scale until a 

weight value is displayed.   

2.  Use a more appropriate unit of 

measure. 

﹍﹍﹍﹍

 

Weight reading below Under load 
limit 

1.  Perform zero calibration 

EEP.E1 

CONFIG parameters are not 
correctly set 

Re-set CONFIG parameters as 
technical manual instructed. 

EEP.E2 

USER parameters are not correctly 
set 

Re-set USER parameters per the 
Technical manual 

CAL.Er 

1.  Input data or loaded weight is too 

small, too big 

2.  Weight signal is unstable, 

un-linear 

1.  Input correct data, load correct 

weight onto platform. 

2.  Return the scale for repair 

Cannot zero the 

display 

1.  Load on scale exceeds allowable 

limits.

20%FS

 

2.  Load on the scale is unstable 

1.  Remove load from the scale. 
2.  Wait for the load to stabilize. then 

press the ZERO/ON/OFF key to 
zero the display   

Weighing is not 

accurate 

1.  An object is stuck between the 

load cell and scale base 

2.  Load cell received a heavy 

impact 

1.  Remove the object 
2. Perform calibration 
3.  Place the load on the center of 

the weighing platform 

 
 
 
 
 

Summary of Contents for 318513

Page 1: ...318513 Operation Manual Contents subject to change without notice Version 1 0 Issue AA ...

Page 2: ......

Page 3: ... 1 2 OVERVIEW OF CONTROLS AND FUNCTIONS 2 Indicator Display Character Definitions 2 Indicator Display 3 Function Keys 3 3 OPERATIONS 4 Normal Weighing Mode 4 ZERO 4 Setting a Tare Weight 4 4 Calibration 5 5 Configuration Parameters Setup 6 6 User Parameters Setup 10 7 Symbol Definitions 14 8 Trouble shooting 15 ...

Page 4: ...llow sufficient warm up time Turn the scale on and allow up to 2 minutes for internal components to stabilize before weighing Electronic scales are precision instruments Do not operate near cell phones radios computers or other electronic devices that emit radio frequencies that may cause unstable readings Avoid using in heavy vibration or heavy airflow conditions Specifications Model 318513 Max C...

Page 5: ...318513 www globalindustrial com 2 2 OVERVIEW OF CONTROLS AND FUNCTIONS Indicator Display Character Definitions ...

Page 6: ...rt 3 seconds Enter or exit HOLD mode Setup or Calibration mode Shift the flashing data entry position from right to left Weighing mode Select weight unit of measure Setup or Calibration mode Increase the digit in the flashing data entry position by 1 Weighing mode Tare the weight Setup or Calibration mode Confirm the input data and continue to next step Weighing mode 3 seconds Zero the platform we...

Page 7: ...wer off the scale by pressing and holding the ZERO ON OFF key for 4 seconds ZERO If the display dose not shows 0 and there is nothing on the platter press the ZERO ON OFF key to zero the reading Zero range 20 full Capacity The zero function is unavailable when the displayed reading is out of the zero range and the indicator will show the error message 0 or 0 meaning the scale is over or under zero...

Page 8: ... the scale becomes stable it will automatically be directed to next step If the second point cannot be calibrated correctly it will display CAL Er and return back to step2 for re calibration 4 When CAL P2 is displayed the scale will be calibrated on third calibration point xxxxxx kg or lb will be displayed The default standard weight is 100 FS Load 10 100 FS this must be equal or larger than the w...

Page 9: ...t unit 100 20000 800 C2 100 25000 Display Resolution for second weight unit if main unit is kg or lb then second unit is lb or kg 100 25000 Note it must be equal or less than 1 25 C1 910 C3 0 1 2 Division select 0 1 1 2 2 5 2 C4 0 5 Decimal point 0 x1 1 x0 1 2 x0 01 3 x0 001 4 x0 0001 5 x10 1 C5 0 1 Main weight unit Calibration and setting capacity weight units 0 kg 1 lb 1 C6 0 1 2 Weight Units th...

Page 10: ... on and current weight signal is NOT within the power on zero point range 0 Prompt power on zero point is over range 1 Current weight 2 Calibration zero point 3 Power off zero point power off tare weight as current tare weight 0 C11 6 Zero tracking range 0 no tracking 1 0 25d 2 0 5d 3 1d 4 1 5d 5 2d 6 3d 7 4d 8 5d 9 6d 6 C12 2 Digital filter intensity 0 very weak 1 weak 2 middle 3 strong 2 C13 5 M...

Page 11: ...05kg 0 005kg 0 01 lb 0 05kg 0 05kg 0 1 lb 0 5kg 0 5kg 1 lb 5kg 5kg 10 lb 50kg 50kg Not available Table2 use LB as calibration unit Calibration division value Display division value in different weight unit that can be used Kg lb 0 0001lb Not available 0 0001lb 0 001 lb 0 0005 kg 0 001 lb 0 01 lb 0 005 kg 0 01 lb 0 1 lb 0 05 kg 0 1 lb 1 lb 0 5 kg 1 lb 10 lb 5 kg 10 lb 0 0002 lb 0 0001 kg 0 0002 lb ...

Page 12: ...ion on C16 If you need to recover all configuration and user parameters to their default value modify the 0 to 1 when C16 0 is shown then press the TARE key to confirm The indicator will display dEF and is flashed press the TARE key to confirm then dEF will flash to indicate recovery processing C16 0 will be displayed when the processing is complete When the indicator displays dEF and is flashed p...

Page 13: ... 2 U3 0 1 2 HOLD PRINT key function set 0 HOLD 1 PRINT 2 HOLD and PRINT 2 U4 0 50 Hold function mode 0 no hold function 1 hold larger weight reading 2 50 when weight is more than 10d and the variety is within 2d 50d hold stable weight When weight is below 10d and then over 10d and becomes stable the new stable weight will be held 2 U5 0 7 Serial communication output format 0 communication is disab...

Page 14: ...current status data it does not receive any data The output format is same as U5 1 5 U5 4 Continuously output of the current gross weight tare weight net weight data weight unit and current status data it does not receive any data The output format is same as U5 2 6 U5 5 When the scale is stable it will output the current displayed reading weight unit and current status data automatically one time...

Page 15: ...larity character including minus sign for negative weight and a space character for positive weight W1 W8 Weight data dp Decimal point U1U2 Measure units kg lb or lb oz d Commands and responses 1 Command W CR 57h 0dh Response over capacity LF u1u2 CR LF H1H2H3 CR ETX under capacity LF _________u1u2 CR LF H1H2H3 CR ETX zero point error LF u1u2 CR LF H1H2H3 CR ETX Note If the weight unit is lb oz U1...

Page 16: ...O ON OFF key for 4 seconds 8 Command all others Response LF CR ETX Unrecognized command e Additional Commands and Responses for Scale Base Application 1 Command F CR 46h 0dh to restore factory calibration data Response LF OK H1H2H3 CR ETX 2 Command O CR 4Fh 0dh zero point calibration Response LF OK H1H2H3 CR ETX if zero calibration is sufficient LF H1H2H3 CR ETX if zero calibration resulted in an ...

Page 17: ...he setting range Ad Analog digital converter chip over max range Ad Analog digital converter chip below min range Weight signal is too large Weight signal is too small EEP E1 Config parameters incorrect no set no calibration over normal range etc EEP E2 User parameters are incorrect CAL Px Calibration point CAL Er Error in calibration CAP The setting full capacity will be displayed Cx y No x confi...

Page 18: ... exceeds overload limit 2 The weight value cannot be displayed in the current unit of measure because it exceeds 6 digits 1 Reduce load on the scale until a weight value is displayed 2 Use a more appropriate unit of measure Weight reading below Under load limit 1 Perform zero calibration EEP E1 CONFIG parameters are not correctly set Re set CONFIG parameters as technical manual instructed EEP E2 U...

Page 19: ...318513 www globalindustrial com 16 ...

Page 20: ...11 Harbor Park Drive Port Washington NY 11050 ...

Reviews: