Model GLI-9143 Portable D.O. Meter

8

Rev. 0-401

BATTERY REPLACEMENT

When the batteries are run
down "LOBAT" is displayed
on the Liquid Crystal Dis-
play.

This is to inform the user that the display will
be shut-off after about 4 hours of use to pre-
vent erroneous measurements due to low
voltage.

Battery replacement
must only take place in
a non-hazardous area
using 1.5V alkaline AA
type batteries. To
replace run-down bat-
teries, simply remove
the two screws on the
rear cover of the in-
strument and replace
the four 1.5V AA bat-
teries with new ones,
observing correct po-
larity.

A 12 VDC power adapter can also be used to
power the unit.

NOTE:

The instrument uses the following
configuration.

It is recommendable to purchase the GLI
voltage adapters that use the proper polarity
configuration. However, the meter can be
used with other adapters. In this case, re-
member to check the correct polarity of your
adapter before connecting it to the meter.

WARRANTY

GLI International, Inc. warrants the Model
GLI-9143 Hand-held/Portable Dissolved Oxy-
gen Meter to be free from defects in material
or workmanship 

for a period of one year (12

months) 

from the date of shipment of this

product from our facility. The GLI-76407/4 or
GLI-76407/10 Dissolved Oxygen Sensor is
warranted 

for a period of 6 months

. A war-

ranty claim will not be honored if defects are
not reported within the warranty period, or if
GLI International determines that defects or
damages are due to normal wear, misappli-
cation, lack of maintenance, abuse, improper
installation, alteration or abnormal conditions.
GLI International’s obligation under this war-
ranty shall be limited to, at its option,
replacement or repair of this product. The
product must be returned to GLI International,
freight prepaid, for examination. The product
must be thoroughly cleaned and any process
chemicals removed before it will be accepted
for replacement or repair. GLI International’s
liability shall not exceed the cost of the prod-
uct. Under no circumstances will GLI
International be liable for any incidental or
consequential damages, whether to person
or property. GLI International will not be liable
for any other loss, damage or expense of any
kind, including loss of profits, resulting from
the installation, use, or inability to use this
product.

Summary of Contents for GLI-9143

Page 1: ...Manual No 9143 Revision 0 401 MODEL GLI 9143 HAND HELD PORTABLE DISSOLVED OXYGEN METER ...

Page 2: ... material until you are sure that the instrument func tions correctly All defective items must be returned in their original packaging together with the supplied accessories GENERAL DESCRIPTION The GLI 9143 is a water resistant micro processor based auto calibration Dissolved Oxygen meter with automatic temperature compensation ATC It has been developed for dissolved oxygen and temperature measure...

Page 3: ... Polypropylene Probe Body 5 Temperature Sensor 6 O ring Seal 7 Silver Chloride Anode sensor element 8 Platinum Cathode sensor element 9 Oxygen permeable Teflon Membrane 10 Protective Cap FUNCTIONAL DESCRIPTION METER 1 Battery compartment 2 Probe connector 3 Liquid crystal display 4 ON OFF button 5 CAL button to enter or exit calibration mode 6 FACTOR button to select altitude F1 and salinity F2 7 ...

Page 4: ...76407A P supplied with the meter and make sure that the rubber O ring sits properly inside the membrane cap 4 Rinse the mem brane cap with electrolyte while shaking it gently Refill with clean electrolyte 5 Gently tap the sides of the mem brane cap with your finger tip to ensure that no air bubbles remain trapped To avoid damaging the membrane do not tap the membrane cap directly on the bottom 6 W...

Page 5: ...ss FACTOR and ensure F1 and F2 are set to the appropriate altitude and salinity values see ALTITUDE and SALINITY COMPENSATION NOTE The instrument must be calibrated whenever the probe membrane or the electrolyte is changed To exit the calibration mode during calibration press CAL The display may be switched from readings in saturation to ppm without re calibration by simply pressing RANGE MEASUREM...

Page 6: ...y will show the measured temperature in degrees Celsius Allow the probe to reach thermal equilibrium with the sample before taking any measure ment The greater the difference between the ambient temperature and the temperature of the sample the longer it will take the probe to equilibrate to the sample ALTITUDE COMPENSATION Press FACTOR and F1 will be displayed Use the UP and the DOWN keys to set ...

Page 7: ...shed or stained which could be due to contact with certain gases or extended use with a loose or damaged mem brane the cathode should be cleaned 1 Use a clean lint free cardboard or cloth and rub the cathode very gently side to side 4 5 times This will be enough to polish and remove any stains without damaging the platinum tip 2 Rinse the probe with deionized or dis tilled water 3 Install a new me...

Page 8: ...Meter to be free from defects in material or workmanship for a period of one year 12 months from the date of shipment of this product from our facility The GLI 76407 4 or GLI 76407 10 Dissolved Oxygen Sensor is warranted for a period of 6 months A war ranty claim will not be honored if defects are not reported within the warranty period or if GLI International determines that defects or damages ar...

Reviews: