background image

Rev. 0-401

7

Model GLI-9143 Portable D.O. Meter

2.

Unscrew the membrane cap by turning it
counterclockwise (see Figure 2).

3.

Wet the sensor by
soaking the bottom 2½
cm (1 inch) of the
probe in electrolyte
(GLI-7041S) for 5 min-
utes.

4.

Rinse a new mem-
brane cap (GLI-76407
A/P) with electrolyte
while shaking it gently.
Refill with clean elec-
trolyte.

5.

Gently tap the sides of
the membrane with
your finger tip to en-
sure that no air
bubbles remain trap-
ped. Do not directly
tap the bottom as this may cause irrepa-
rable damage to the membrane.

6.

Make sure that the rubber O-ring is
seated properly inside the membrane
cap.

7.

With the sensor facing down, screw the
membrane cap clockwise. Some elec-
trolyte will overflow.

CLEANING PLATINUM CATHODE

The platinum cathode (item # 8 in the
FUNCTIONAL DESCRIPTION - PROBE on
page 3) should always be bright and untar-
nished. If it is tarnished or stained, which

could be due to contact with certain gases or
extended use with a loose or damaged mem-
brane, the cathode should be cleaned:

1.

Use a clean, lint-free cardboard or cloth
and rub the cathode very gently side to
side 4-5 times. This will be enough to
polish and remove any stains without
damaging the platinum tip.

2.

Rinse the probe with deionized or dis-
tilled water.

3.

Install a new membrane cap and fill it
with fresh electrolyte following the previ-
ous instructions.

4.

Recalibrate the instrument.

CLEANING MEMBRANE

In order to have accurate and stable meas-
urements, it is important that the surface of
the membrane is in perfect condition. This
permeable membrane isolates the sensor
elements from the environment but allows
oxygen to enter. If any dirt is deposited on the
membrane:

1.

Rinse it carefully with distilled or deion-
ized water.

2.

If any imperfections are observed, or any
damage is evident (such as wrinkles,
tears or holes), the membrane cap
should be replaced.

3.

Make sure that the O-ring is properly
seated in the membrane cap.

FIGURE 2

Summary of Contents for GLI-9143

Page 1: ...Manual No 9143 Revision 0 401 MODEL GLI 9143 HAND HELD PORTABLE DISSOLVED OXYGEN METER ...

Page 2: ... material until you are sure that the instrument func tions correctly All defective items must be returned in their original packaging together with the supplied accessories GENERAL DESCRIPTION The GLI 9143 is a water resistant micro processor based auto calibration Dissolved Oxygen meter with automatic temperature compensation ATC It has been developed for dissolved oxygen and temperature measure...

Page 3: ... Polypropylene Probe Body 5 Temperature Sensor 6 O ring Seal 7 Silver Chloride Anode sensor element 8 Platinum Cathode sensor element 9 Oxygen permeable Teflon Membrane 10 Protective Cap FUNCTIONAL DESCRIPTION METER 1 Battery compartment 2 Probe connector 3 Liquid crystal display 4 ON OFF button 5 CAL button to enter or exit calibration mode 6 FACTOR button to select altitude F1 and salinity F2 7 ...

Page 4: ...76407A P supplied with the meter and make sure that the rubber O ring sits properly inside the membrane cap 4 Rinse the mem brane cap with electrolyte while shaking it gently Refill with clean electrolyte 5 Gently tap the sides of the mem brane cap with your finger tip to ensure that no air bubbles remain trapped To avoid damaging the membrane do not tap the membrane cap directly on the bottom 6 W...

Page 5: ...ss FACTOR and ensure F1 and F2 are set to the appropriate altitude and salinity values see ALTITUDE and SALINITY COMPENSATION NOTE The instrument must be calibrated whenever the probe membrane or the electrolyte is changed To exit the calibration mode during calibration press CAL The display may be switched from readings in saturation to ppm without re calibration by simply pressing RANGE MEASUREM...

Page 6: ...y will show the measured temperature in degrees Celsius Allow the probe to reach thermal equilibrium with the sample before taking any measure ment The greater the difference between the ambient temperature and the temperature of the sample the longer it will take the probe to equilibrate to the sample ALTITUDE COMPENSATION Press FACTOR and F1 will be displayed Use the UP and the DOWN keys to set ...

Page 7: ...shed or stained which could be due to contact with certain gases or extended use with a loose or damaged mem brane the cathode should be cleaned 1 Use a clean lint free cardboard or cloth and rub the cathode very gently side to side 4 5 times This will be enough to polish and remove any stains without damaging the platinum tip 2 Rinse the probe with deionized or dis tilled water 3 Install a new me...

Page 8: ...Meter to be free from defects in material or workmanship for a period of one year 12 months from the date of shipment of this product from our facility The GLI 76407 4 or GLI 76407 10 Dissolved Oxygen Sensor is warranted for a period of 6 months A war ranty claim will not be honored if defects are not reported within the warranty period or if GLI International determines that defects or damages ar...

Reviews: