background image

Model GLI-9143 Portable D.O. Meter

4

Rev. 0-401

SPECIFICATIONS

Range

D.O.

% Sat.

Temp.

0.00 to 45.00 ppm
0.0 to 300.0%
0.0 to 50.0°C

Resolution

D.O.

% Sat.

Temp.

0.01 ppm
0.1%
0.1°C

Accuracy

D.O.

(@25°C/77°F) % Sat.

Temp.

± 1.5% full scale ppm
± 1.5% full scale %
± 0.5°C

Typical EMC

D.O.

Deviation

% Sat.

Temp.

± 0.3 ppm
± 3.5%
± 0.5°C

Calibration

Automatic in saturated air

Temp. Compensation

Automatic from 0 to 50°C
(32 to 122

°

F)

Altitude Compensation

0-13,123 ft. (0-4000 m) in
328 ft. (100 m) increments

Salinity Compensation

0-80 g/l in 1 g/l increments

Operating Conditions

From 0 to 50

°

C (32 to

122

°

F); 100% relative

humidity

Power Requirement

Four 1.5 volt AA alkaline
type batteries; 200 hours
continuous use; auto shut-
off after 4 hours of non-
use; power socket for 12
VDC adapter

Dimensions

7.7 x 3.1 x 2.4 inches
(196 x 80 x 60 mm)

Weight

15 oz. (425 g)

PROBE PREPARATION

All probes are shipped dry. To hydrate the
probe and prepare it for use, connect it to the
meter and proceed as follows:

1.

Remove the red and black
plastic shipping cap which can
be discarded.

2

Wet the sensor by soaking the
bottom 2½ cm (1 inch) of the
probe in electrolyte (GLI-7041S) for 5
minutes.

3

Take a membrane cap (GLI-76407A/P
supplied with the meter) and make sure
that the rubber O-ring sits properly inside
the membrane cap.

4

Rinse the mem-
brane cap with
electrolyte while
shaking it gently.
Refill with clean
electrolyte.

5

Gently tap the
sides of the mem-
brane cap with
your finger tip to
ensure that no air
bubbles remain
trapped. To avoid
damaging the membrane, do not tap the
membrane cap directly on the bottom.

6

With the sensor facing down, screw the
membrane cap clockwise. Some elec-
trolyte will over-flow.

When not in use, protect the mem-
brane by placing the protective cap
on it.

CALIBRATION

PROBE POLARIZATION

The probe is under polarization with a fixed
voltage of approximately 800 mV. Probe po-
larization is essential for stable meas-
urements with the same recurring degree of
accuracy.

With the probe properly polarized, oxygen is
continually "consumed" by passing through
the sensitive membrane and dissolving in the
electrolyte solution contained inside the
probe. If this operation is interrupted, the
electrolyte solution continues to be enriched
with oxygen until it reaches equilibrium with
the surrounding solution. Whenever meas-
urements are taken with a non-polarized
probe, the oxygen level indicated is that of
the test solution as well as any oxygen pres-
ent in the electrolyte solution. This reading is
obviously incorrect. The GLI meter automati-

Summary of Contents for GLI-9143

Page 1: ...Manual No 9143 Revision 0 401 MODEL GLI 9143 HAND HELD PORTABLE DISSOLVED OXYGEN METER ...

Page 2: ... material until you are sure that the instrument func tions correctly All defective items must be returned in their original packaging together with the supplied accessories GENERAL DESCRIPTION The GLI 9143 is a water resistant micro processor based auto calibration Dissolved Oxygen meter with automatic temperature compensation ATC It has been developed for dissolved oxygen and temperature measure...

Page 3: ... Polypropylene Probe Body 5 Temperature Sensor 6 O ring Seal 7 Silver Chloride Anode sensor element 8 Platinum Cathode sensor element 9 Oxygen permeable Teflon Membrane 10 Protective Cap FUNCTIONAL DESCRIPTION METER 1 Battery compartment 2 Probe connector 3 Liquid crystal display 4 ON OFF button 5 CAL button to enter or exit calibration mode 6 FACTOR button to select altitude F1 and salinity F2 7 ...

Page 4: ...76407A P supplied with the meter and make sure that the rubber O ring sits properly inside the membrane cap 4 Rinse the mem brane cap with electrolyte while shaking it gently Refill with clean electrolyte 5 Gently tap the sides of the mem brane cap with your finger tip to ensure that no air bubbles remain trapped To avoid damaging the membrane do not tap the membrane cap directly on the bottom 6 W...

Page 5: ...ss FACTOR and ensure F1 and F2 are set to the appropriate altitude and salinity values see ALTITUDE and SALINITY COMPENSATION NOTE The instrument must be calibrated whenever the probe membrane or the electrolyte is changed To exit the calibration mode during calibration press CAL The display may be switched from readings in saturation to ppm without re calibration by simply pressing RANGE MEASUREM...

Page 6: ...y will show the measured temperature in degrees Celsius Allow the probe to reach thermal equilibrium with the sample before taking any measure ment The greater the difference between the ambient temperature and the temperature of the sample the longer it will take the probe to equilibrate to the sample ALTITUDE COMPENSATION Press FACTOR and F1 will be displayed Use the UP and the DOWN keys to set ...

Page 7: ...shed or stained which could be due to contact with certain gases or extended use with a loose or damaged mem brane the cathode should be cleaned 1 Use a clean lint free cardboard or cloth and rub the cathode very gently side to side 4 5 times This will be enough to polish and remove any stains without damaging the platinum tip 2 Rinse the probe with deionized or dis tilled water 3 Install a new me...

Page 8: ...Meter to be free from defects in material or workmanship for a period of one year 12 months from the date of shipment of this product from our facility The GLI 76407 4 or GLI 76407 10 Dissolved Oxygen Sensor is warranted for a period of 6 months A war ranty claim will not be honored if defects are not reported within the warranty period or if GLI International determines that defects or damages ar...

Reviews: