background image

Rev. 0-401

5

Model GLI-9143 Portable D.O. Meter

cally polarizes the probe when it is switched
on.

Calibration is simple and is recommended
every time the meter is switched on.

1.

Make sure the probe is ready for meas-
urement (see PROBE PREPARATION),
i.e. the membrane is filled with electrolyte
and the probe is connected to the meter
and properly polarized.

2.

Switch the meter on.

3.

"COND" appears on the
display to inform you
that the probe is in auto-
conditioning (automatic
polarization) mode.

4.

Once "COND" disappears, the probe is
polarized and the instrument can be cali-
brated.

5.

For an accurate calibration, it is recom-
mended to wait an additional
5 or 10 minutes to ensure
optimum conditioning of the
probe.

6.

Remove the protective cap.

7.

Press CAL. "----" appears
on the larger LCD to-
gether with "CAL" to
indicate that the instru-
ment is in calibration mode.

8.

The instrument will
automatically standard-
ize itself to the actual
saturation value. After
approximately 1 minute,
it will show "100%" on
the LCD and a small
"SAMPLE" to indicate
that the calibration is complete.

9.

Press FACTOR and ensure F1 and F2
are set to the appropriate altitude and

salinity values (see ALTITUDE and
SALINITY COMPENSATION).

NOTE:

The instrument must be calibrated
whenever the probe, membrane or
the electrolyte is changed.

To exit the calibration mode
during calibration, press CAL.

The display may be switched
from readings in % saturation
to ppm without re-calibration,
by simply pressing RANGE.

MEASUREMENT

Make sure the meter has been cali-
brated and the protective cap has
been removed. Immerse the tip of
the probe in the sample to be tested
and ensure that the temperature
sensor is also immersed.

To display values in % satura-
tion, press RANGE.

For accurate dissolved oxygen measure-
ments, a water movement of at least 30 cm
(12 inches)/sec is required. This is to ensure
that oxygen depleted on the membrane sur-
face is constantly replenished. A moving
stream will provide adequate circulation.

During field measurements, this condition
may be obtained by manually stirring the
probe. Accurate readings are not possible
while the liquid is stationary.

During laboratory measurements, the use of
a magnetic stirrer to ensure a certain agita-
tion of the fluid is recommended. This way,
any errors due to the presence of air bubbles
on the membrane surface are minimized.

Summary of Contents for GLI-9143

Page 1: ...Manual No 9143 Revision 0 401 MODEL GLI 9143 HAND HELD PORTABLE DISSOLVED OXYGEN METER ...

Page 2: ... material until you are sure that the instrument func tions correctly All defective items must be returned in their original packaging together with the supplied accessories GENERAL DESCRIPTION The GLI 9143 is a water resistant micro processor based auto calibration Dissolved Oxygen meter with automatic temperature compensation ATC It has been developed for dissolved oxygen and temperature measure...

Page 3: ... Polypropylene Probe Body 5 Temperature Sensor 6 O ring Seal 7 Silver Chloride Anode sensor element 8 Platinum Cathode sensor element 9 Oxygen permeable Teflon Membrane 10 Protective Cap FUNCTIONAL DESCRIPTION METER 1 Battery compartment 2 Probe connector 3 Liquid crystal display 4 ON OFF button 5 CAL button to enter or exit calibration mode 6 FACTOR button to select altitude F1 and salinity F2 7 ...

Page 4: ...76407A P supplied with the meter and make sure that the rubber O ring sits properly inside the membrane cap 4 Rinse the mem brane cap with electrolyte while shaking it gently Refill with clean electrolyte 5 Gently tap the sides of the mem brane cap with your finger tip to ensure that no air bubbles remain trapped To avoid damaging the membrane do not tap the membrane cap directly on the bottom 6 W...

Page 5: ...ss FACTOR and ensure F1 and F2 are set to the appropriate altitude and salinity values see ALTITUDE and SALINITY COMPENSATION NOTE The instrument must be calibrated whenever the probe membrane or the electrolyte is changed To exit the calibration mode during calibration press CAL The display may be switched from readings in saturation to ppm without re calibration by simply pressing RANGE MEASUREM...

Page 6: ...y will show the measured temperature in degrees Celsius Allow the probe to reach thermal equilibrium with the sample before taking any measure ment The greater the difference between the ambient temperature and the temperature of the sample the longer it will take the probe to equilibrate to the sample ALTITUDE COMPENSATION Press FACTOR and F1 will be displayed Use the UP and the DOWN keys to set ...

Page 7: ...shed or stained which could be due to contact with certain gases or extended use with a loose or damaged mem brane the cathode should be cleaned 1 Use a clean lint free cardboard or cloth and rub the cathode very gently side to side 4 5 times This will be enough to polish and remove any stains without damaging the platinum tip 2 Rinse the probe with deionized or dis tilled water 3 Install a new me...

Page 8: ...Meter to be free from defects in material or workmanship for a period of one year 12 months from the date of shipment of this product from our facility The GLI 76407 4 or GLI 76407 10 Dissolved Oxygen Sensor is warranted for a period of 6 months A war ranty claim will not be honored if defects are not reported within the warranty period or if GLI International determines that defects or damages ar...

Reviews: