Glas-Col Temp-O-Control TOT-1200 Operating Instructions Manual Download Page 7

 

 
 
 
 
 

 

DESCRIPTION OF AUTO TUNING FUNCTION 

 
 
Auto tuning is an internal algorithm that allows your controller to learn the characteristics 
of your set-up. It will determine the right amount of power to apply to your heating 
device to bring your process to the set point temperature in the least amount of time 
without overshooting and will regulate the temperature as tightly as possible. It is 
important for you to understand the mechanism of Proportional, Integral, and Derivative 
Control in order to understand the instrument’s ability to control your process. 
 
To auto tune the system you press the advance button one time and AUT will appear on 
display. Use arrow buttons and change to YES and then press the infinity button. The 
instrument immediately goes into the AUTO TUNE mode and will run for 30 min. to 90 
minutes. When the AUTO TUNE mode is finished it will stop flashing between the auto 
tune screen and operation screen. 
 

 

* Since ON-OFF action (two-position action) takes place in the auto tune procedure, the 
temperature of the process material being heated may change greatly. Do not carry out 
the auto tune if the process does not allow a significant variation of the process 
temperature. Use a material that is similar but expendable for the auto tune procedure. 
* Re-run the auto tune again if the Set-point Value temperature (SV) is significantly 
changed, the upper or lower PV temperature has changed, or if the controlled equipment 
is changed. 
 

 
 

DESCRIPTION OF PID CONTROL 

 

“PID CONTROL”

 stands for Proportional with Integral and Derivative control. It is a 

mathematical method of predicting how much power must be applied to the system to 
reach the Set-point temperature with a minimum of temperature oscillation (overshooting 
and undershooting the set point temperature.) A brief description of each component of 
PID Control is discussed below: 
 

Proportional Control Band: 

The proportional band is that area around the set-point 

value temperature where the power supplied to the output is neither fully ON nor fully 
OFF. The proportional band can be moved to higher or lower temperatures by the integral 
and derivative control parameters. For ON/OFF control, set the proportional band to “0”. 

Summary of Contents for Temp-O-Control TOT-1200

Page 1: ...TEMP O TROL OPERATING INSTRUCTIONS MODELS TOT 1200 TOT 1200A...

Page 2: ...the specified maximum total electrical loading for this instrument See specifications Avoid installing the instruments where it is likely to come into contact with steam or condensing vapors Avoid in...

Page 3: ...he unit 2 Controller Allows user to program the temperature set point variables and system alarms and to monitor the Process temperature 3 TC Jack This is a color coded miniature thermocouple panel ja...

Page 4: ...with a three prong grounded plug for use in standard 120 Volt 15 amp wall receptacles CONTROLLER MODULE DESCRIPTION The illustration above is of the front panel controller module on the Model TOT 1200...

Page 5: ...out the warranty registration card and return it to us Should anything appear to be missing or the unit appears to be damaged contact us immediately Take the time to familiarize yourself with the con...

Page 6: ...e process temperature PT as read by the thermocouple will appear on the display in red and the set point temperature will appear on the right side of the display in green Congratulations you are now r...

Page 7: ...une procedure the temperature of the process material being heated may change greatly Do not carry out the auto tune if the process does not allow a significant variation of the process temperature Us...

Page 8: ...ime value should eliminate the offset without overcompensating leading to oscillation Derivative Time rate The derivative time is that time used in calculating the rate of change and the thermal lag T...

Page 9: ...9 Control Menu Settings...

Page 10: ...10...

Page 11: ...11...

Page 12: ...12...

Page 13: ...ve the problem INACCURATE READINGS Your instrument is or was factory calibrated to our stated specifications Field re calibration is not possible since this is a software re calibration The accuracy o...

Page 14: ...oper installation or through attempts to operate the apparatus beyond its rated capacity intentional or otherwise Normally expendable parts are not covered by this warranty LIMITATION OF WARRANTIES AP...

Page 15: ...g the product Reference the RGA number on the shipping box and on a written description of the problem A 15 restocking charge of the net price is changed for all standard products returned to stock Gl...

Reviews: