5
3. Alarm Setpoint - This control will illuminate the number 2 on the display when
the alarm is triggered. This is controlled by the A.Hi and A.Lo
setting in operations menu.
4. Enter Key -
All menu items and program variables are selected and
entered using this key with either the up or down arrow key.
5. Down Arrow -
Down arrow will either step down in a menu or decrease the
program variable when used with the enter key.
6. Up Arrow -
Up arrow will either set up in a menu or increase the program
variable when used with the enter key.
INSTALLATION
When you receive your instrument, carefully unpack and inspect your instrument
for damage and missing hardware. Take the time to fill out the warranty registration card
and return it to us. Should anything appear to be missing or the unit appears to be
damaged, contact us immediately.
Take the time to familiarize yourself with the controls and inputs to the
instrument. The instrument can be mounted on lab rods and ring stands using the
mounting clamp on the rear of the unit. Mount the unit securely before connecting the
unit to the AC line source.
Warning!
This instrument contains hazardous operating voltages. Disconnect the unit from the
main power supply and the inputs when installing to avoid electrical shock.
Note
The absolute maximum output power is 1200 Watts resistive.
Using a variac with this instrument may permanently damage the unit.
Caution!
It is always recommended that every lab temperature controller be used with an
independent high-temperature cut-off device such as an Over-Temp Probe.