MA531_EN
EN 36
UTC2
WORKING SCREEN
(Ultrasuond)
P1 10:00
0.0 W
0.0 W/cm
2
1 MHz
100%
Treatment in
START
Press the key START to begin the treatment. In START condition, warned by the colour of P1 that change in yellow and
by the writing “Treatment in START”, select the icone effective intensity (Watt/cm
2
), the ultrasound level will be regulated
(the treatment time has calculated only when the output power is greater than 0.0 W).
Effective
Intensity
Frequency
Modulation
0.1 - 2.0 / 3.0
1, 2, 3 MHz
5-100%
In the case of active control contact (see “Contact Control” in Settings) must be push START and the handle must be well
in contact with the area to treat.
Note: Increase slowly the output level so that the patient warns a light sense of heat without to feel any bother. It is prefer-
able to move the handpiece circularly or with “coming and going” slow and uniform.
Pressing the key Stop (START/STOP) it is possible to interrupt the treatment. The timer and the power from the two out-
puts are stopped. To start again the treatment press the key START (START/STOP). To exit from the mask of work press
the key HOME.
When the treatment time has elapsed the equipment’s screen give the message “Ended Treatment”, together with an